Uncertainty evaluation of the static expansion system and its long-term stability at NMIJ
Source abstract
Abstract
Abstract A pressure standard based on the static expansion of gases has been made available at the National Metrology Institute of Japan (NMIJ) for gauge calibrations in the range of 0.1 mPa–2 kPa over the past 30 years. The static expansion system (SES) generates calibration pressures by multiple-step gas expansions using two expansion ratios ( R 900 ≃907.3 and R 50 ≃53.4). These expansion ratios were determined using the two reference gauges method instead of the one reference gauge method, which has been in use thus far. Relative uncertainties of 0.073% (k = 1) for R 900 and 0.036% (k = 1) for R 50 were obtained, and the calibration uncertainties were slightly improved compared to the current calibration and measurement capability (CMC) of the SES at NMIJ as follows: U = 0.28% (k = 2) (1 mPa–10 Pa) for spinning rotor gauges (SRGs). U = 0.23% (k = 2) (0.2 Pa–10 Pa) for capacitance diaphragm gauges (CDGs). U = 0.17% (k = 2) (10 Pa–110 Pa) for CDGs. In addition, the calibration results of the check standards of the past 12 years are reported. The data suggest that the SES was stable compared to the calibration uncertainty over the years.
Source: openaire · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Scientific Measurement and Uncertainty Evaluation · Advanced Electrical Measurement Techniques · Scientific Research and Discoveries
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (34)
Show source-provided references
1. Discussion of methods for the estimation of volumetric ratios determined by multiple expansions10.1016/S0042-207X(98)00350-9 ↗
2. The Reference Gauge technique for static expansion ratios-Applied to NPL medium vacuum standard SEA310.1016/j.vacuum.2006.06.012 ↗
3. Volume ratio determination in static expansion systems by means of a spinning rotor gauge10.1016/S0042-207X(98)00337-6 ↗
4. High-accuracy calibration in the vacuum range 0.3 Pa to 4000 Pa using the primary standard of static gas expansion10.1088/0026-1394/39/3/2 ↗
5. Preliminary experiments of vacuum gauge calibration by means of expansion method10.3131/jvsj.32.288 ↗
6. Computer control of an expansion apparatus for gauge calibration10.3131/jvsj.38.278 ↗
7. Results of bilateral pressure comparisons in the vacuum region between the ETL and the IMGC10.1088/0026-1394/34/5/6 ↗
8. Uncertainty evaluation of the IMGC-CNR static expansion system10.1088/0026-1394/41/4/009 ↗
9. The static expansion system as a new medium vacuum primary standard in KRISS10.1088/0026-1394/42/6/S05 ↗
10. The improvement of the volume ratio measurement method in static expansion vacuum system10.1016/j.phpro.2012.03.591 ↗
11. Establishment of Thailand's national primary vacuum standard by a static expansion method10.1088/1742-6596/1380/1/012003 ↗
12. Static expansion primary vacuum standard - Part 1: determination of the volume ratio of the expansion stage
13. Realization of the medium and high vacuum primary standard in CENAM, Mexico10.1088/0026-1394/42/6/S01 ↗
14. The new UME primary standard for pressure generation in the range from 9 × 10-4 Pa to 103 Pa10.1088/0026-1394/41/4/005 ↗
15. Development and characterization of volume expansion vacuum standard
16. The determination of volume ratios by gas depletion through multiple expansions10.1016/j.vacuum.2006.06.014 ↗
17. Measuring volume ratios of vacuum vessels using non-evaporable getters10.1016/j.vacuum.2012.11.010 ↗
18. Final report on the key comparison, CCM.P-K15 in the pressure range from 1.0 × 10-4 Pa to 1.0 Pa
19. Final report on the key comparison CCM.P-K4
20. Yokogawa Electric Corporation
21. Bipm
22. Yokogawa Electric Corporation
23. Characterization of differential pressure gauges focusing on calibration curve and line pressure dependence
24. Spinning rotor vacuum gauges10.1016/0042-207X(82)94048-9 ↗
25. Virial coefficients of pure gases and mixtures
26. Report of pilot study CCM.P-P1 for international comparison of absolute pressure measurements in gas from 3 × 10 -9 Pa to 9 × 10 -4 Pa10.1088/0026-1394/52/1A/07012 ↗
27. Thermal transpiration effect of
28. Compensation of thermal transpiration effect for pressure measurements by capacitance diaphragm gauge10.3131/jvsj2.53.686 ↗
29. Factors affecting the reproducibility of the accommodation coefficient of the spinning rotor gauge10.1116/1.2790910 ↗
30. Spinning rotor viscosity gauge: a transfer standard for the laboratory or an accurate gauge for vacuum process control10.1116/1.571592 ↗
31. Present status of pressure standards in vacuum region10.3131/jvsj.27.747 ↗
32. On the stability of capacitance-diaphragm gauges with ceramic membranes10.1116/1.3529023 ↗
33. Measurement performance of high-accuracy low-pressure transducers
34. Recommended practices for the calibration and use of capacitance diaphragm gages as transfer standards10.1116/1.577209 ↗