Investigation of an extractor gauge modified by carbon nanotubes emitter grown on stainless steel substrate
Source abstract
Abstract
Abstract This paper presents an investigation of carbon nanotube (CNT) emitter-based ionization gauge. The metrological performances of this CNT emitter-based ionization gauge were investigated and studied by numerical and experimental methods. The electron source was constructed with multi-wall nanotubes (MWNT), which were grown on stainless steel substrate using thermal chemical vapor deposition (CVD) technique. The influence of anode and reflector voltages on the anode current Ia and ion current I+ were studied in this work, respectively. The electron transmittance increases significantly along with the increase of anode voltage, and I+ increases dramatically as the increase of reflector voltage from 0 V to 525 V, but it decreases again with the further increase of reflector voltage to 600 V. The studied gauge with CNT cathode has good measurement linearity from 1.0 × 10−8 Pa to 5.0 × 10−4 Pa and a gauge sensitivity of 0.03 Pa−1 for argon is achieved.
Source: openaire · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Carbon Nanotubes in Composites · Analytical Chemistry and Sensors · Electrochemical Analysis and Applications
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (21)
Show source-provided references
1. Redhead · 1968
2. Johnson · 1974
3. Comparison of hot cathode and cold cathode ionization gauges10.1116/1.577439 ↗
4. Starting delays in cold-cathode gauges at low pressures10.1116/1.579944 ↗
5. The development of vacuum measurements down to extremely high vacuum-XHV10.1016/j.measurement.2012.08.018 ↗
6. Field emitter-based vacuum sensors10.1016/j.vacuum.2011.08.008 ↗
7. Space Research & Planetary Sciences Division at the University of Bern, Swiss, Kathrin Altwegg; 2011. http://space.unibe.ch/en/rosina/cops/pictures.html.
8. Ptolemy–An instrument to measure stable isotopic ratios of key volatiles on a cometary nucleus10.1007/s11214-006-9001-5 ↗
9. A vacuum sensor using field emitters made by multiwalled carbon nanotube yarns10.1016/j.vacuum.2011.04.027 ↗
10. Ionization vacuum gauge with a carbon nanotube field electron emitter combined with a shield electrode10.1016/j.vacuum.2009.06.016 ↗
11. Dong · 2003
12. Dahl · 2008
13. Electron-impact ionization cross-sections and ionization rate coefficients for atoms and ions10.1086/190154 ↗
14. The efficiencies of ionization and ionization potentials of various gases under electron impact10.1103/PhysRev.39.270 ↗
15. Vacuum-calibration apparatus with pressure down to 10-10 Pa10.1116/1.3457934 ↗
16. Unique characteristics of cold cathode carbon- nanotube- matrix field emitters10.1103/PhysRevB.55.9391 ↗
17. Field emission from well-aligned, patterned, carbon nanotube emitters10.1063/1.126164 ↗
18. Influences of the surface reactions on the field emission from multiwall carbon nanotubes10.1063/1.1591246 ↗
19. Generalized Fowler-Nordheim theory of field emission of carbon nanotubes10.1103/PhysRevLett.101.027602 ↗
20. Ion deflector of an ionization gauge for extreme high vacuum10.1116/1.578709 ↗
21. Ionization gauge errors in low pressures10.1116/1.581912 ↗