journal article·2019-06

A new high precision, broad range gas micro-flow calibration apparatus

Meiru Guo, Zhengyi Ren, Yongjun Cheng, Wenjun Sun, Yali Li, Lan Zhao, Meng Dong

Vacuum· Volume 164
Publisher page

Abstract

Abstract In this paper, a new calibration apparatus has been developed for gas micro-flow calibration. The apparatus employs both the constant volume method and the constant conductance method. The calibration range from 10−1 to 10−13 Pa·m3/s, and the combined standard uncertainty is 1.2%∼1.8%. When calibrating extremely low gas flow rate, the non-evaporable getter pump is used to reduce the effect of chamber wall outgassing on the measurement results, so that the lower limits of the constant volume method and constant conductance method are extended to 10−10 Pa·m3/s and 10−13 Pa·m3/s, respectively. Then a method for measurement of orifice conductance is proposed, the on-line precise measurement of the orifice conductance is realized; the consistency of flow measurement results by constant volume and constant conductance methods is better than 1.3%, then compared with the same standard leak was calibrated in both constant conductance method of the calibration apparatus and the constant pressure gas micro-flow calibration apparatus of the laboratory, and the deviation of those results was less than 1%.
Source: openaire · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Advanced Sensor Technologies Research · Microfluidic and Capillary Electrophoresis Applications · Flow Measurement and Analysis

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (24)

Show source-provided references
1. A gas flow standard apparatus
2. Ion pump design for improved pumping speed at low pressure10.5757/ASCT.2016.25.6.108 ↗
3. Vacuum-calibration apparatus with pressure down to 10(-10) Pa10.1116/1.3457934 ↗
4. Precision gas flowmeter for vacuum metrology10.1016/0042-207X(93)90362-E ↗
5. Critical review: recommended practices for the calibration and use of leaks10.1116/1.578137 ↗
6. Accurate, wide range ultrahigh vacuum calibration system10.1116/1.1315866 ↗
7. A review of micropumps10.1088/0960-1317/14/6/R01 ↗
8. The national bureau of standards primary high vacuum standard
9. Fundamental leak calibration system for gas leaks with a defined pressure difference over the leak element10.1116/1.579936 ↗
10. American Society for Testing and Materials, Standard Practices for Calibrating Gaseous Reference Leaks, ASTM E908-1998, 23-24.
11. Monte Carlo study of a new PTB primary standard for very low pressures10.1088/0026-1394/36/6/15 ↗
12. The gas flux distribution in the XHV chamber of the vacuum primary standard developed by PTB10.1016/j.vacuum.2003.12.010 ↗
13. Experimental verification of a Monte Carlo simulation of the gas in a vacuum chamber10.1016/j.vacuum.2006.03.011 ↗
14. Gas flow measurements connected with the continuous expansion system
15. Development of vacuum metrology in national institute of metrology
16. Partial pressure measurement standard for characterizing partial pressure analyzers and measuring outgassing rates10.1116/1.4935432 ↗
17. Volume ratio determination in static expansion systems by means of a spinning rotor gauge10.1016/S0042-207X(98)00337-6 ↗
18. Potter · 2001
19. Causes of natural gas pipeline explosive ruptures10.1061/(ASCE)PS.1949-1204.0000116 ↗
20. Numerical simulation of rarefied gas flow through a thin orifice10.1017/S0022112004000710 ↗
21. Transient flow of rarefied gas through an orifice10.1116/1.3682370 ↗
22. Transient flow of rarefied gas through a short tube10.1016/j.vacuum.2012.09.021 ↗
23. Numerical study of unsteady rarefied gas flow through an orifice10.1016/j.vacuum.2014.05.004 ↗
24. ISO/IEC Guide 98-3, Uncertainty of Measurement-Part 3: Guide to the Expression of Uncertainty in Measurement (GUM: 1995).