journal article·2019-06Open access

3D-Simulation of a Bayard Alpert ionisation gauge using SIMION program

Ricardo Silva, Nenad Bundaleski, Ana L. Fonseca, Orlando M.N.D. Teodoro

Vacuum· Volume 164

Abstract

Numerical simulation of a Bayard-Alpert ionisation gauge using SIMION 8.1 software is presented. Gauge sensitivity, being the most important parameter of an ionisation gauge, has been derived using two different approaches. In the first one, SIMION is used to determine mean length of electron trajectories, and ion collection efficiency. From these values, known ionisation cross sections and the gas temperature, the sensitivity can be readily calculated. In the second and more detailed approach, a Monte Carlo method was implemented in the Lua programming language to simulate electron impact ionisation of the gas. The Lua code is incorporated in the SIMION trajectory simulations, in which the ionisation event is modelled by transforming a primary electron into an ion. This approach evaluates the contribution of each trajectory to the gas ionisation. It considers energy dependence of the ionisation probability and the contribution of the electrons backscattered from the grid to the sensitivity. The simulation approaches were tested on the gauge geometry, recently modelled by the CERN group. Very good quantitative agreement with the CERN simulations was obtained using both approaches. Monte Carlo simulations reveal that the relative contribution of electrons backscattered from molybdenum grid to the sensitivity is about 14%.
Source: openalex_inverted_index · Quality: baseline

Research context

OpenAlex topics: Atomic and Molecular Physics · Nuclear Physics and Applications · Electron and X-Ray Spectroscopy Techniques

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (19)

Show source-provided references
1. Sensitivity of hot cathode ionization gauges10.1116/1.572991 ↗
2. Nonstable behaviour of widely used ionization gauges10.1116/1.578835 ↗
3. Causes of nonstability and nonreproducibility in widely used Bayard-Alpert ionisation gauges10.1116/1.578836 ↗
4. Sensitivity variations in the Bayard-Alpert ionisation gauge, Part II: simulations of ion collection, for collimated beam input and chaotic gas input, using a digital computer10.1016/0042-207X(68)92089-7 ↗
5. Dependence of sensitivity coefficient of a nude-type Bayard-Alpert gauge on the diameter of an envelope10.1016/S0042-207X(98)00353-4 ↗
6. New ionization gauge geometries providing stable and reproducible sensitivities10.1116/1.572714 ↗
7. Stable and reproducible Bayard-Alpert ionization gauge10.1116/1.578837 ↗
8. Numerical investigations of hot cathode ionisation gauges10.1016/S0042-207X(98)00133-X ↗
9. 3D-simulation of ionisation gauges and comparison with measurements10.1016/j.vacuum.2016.12.014 ↗
10. The effect of ambient temperature on the sensitivity of hot-cathode ionization gauges10.1016/j.vacuum.2004.09.013 ↗
11. New model for electron-impact ionization cross sections of molecules10.1063/1.471116 ↗
12. Recommended data on the electron impact ionisation of atoms and ions: fluorine to nickel10.1063/1.555809 ↗
13. Secondary electron emission in the scanning electron microscope10.1063/1.332840 ↗
14. CASINO V2.42—a fast and easy-to-use modeling tool for scanning electron microscopy and microanalysis users10.1002/sca.20000 ↗
15. Theory experiment comparison of the electron backscattering factor, from solids at low electron energy (250–5,000 eV)10.1002/sca.20091 ↗
16. Measuring the angular dependent energy distribution of backscattered electrons at variable geometry10.1002/sca.4950270605 ↗
17. Absolute partial and total cross sections for electron-impact ionization of argon from threshold to 1000 eV10.1103/PhysRevA.52.1115 ↗
18. An improved ion gauge with gold coated electrodes for reliable operation in reactive gases and for use as reference standard10.1016/0042-207X(85)90005-3 ↗
19. Secondary electrons in ion gauges10.1116/1.574846 ↗