article·2015-01-23

Calibrating the Z-magnification of atomic force microscope below 10 nm by single-atom steps

Bo-Ching He, Wei-En Fu, Chung-Lin Wu, Yun-San Chien, Huay‐Chung Liou

Publisher page

Abstract

Abstract This study identifies a procedure for the calibration of the Z-magnification of an atomic force microscope below 10 nm by single-atom steps and gives an uncertainty evaluation. A Si(111) surface with periodically arranged single-atom steps was used as a standard reference. The lattice constant of the Si(111) is 0.312 nm, which is traced to SI units through X-ray diffraction. The 33 layers of continuous stacked single-atom steps were scanned and step heights ranging from 0.3 nm to 10 nm were measured and analyzed. The influential factors of calibration for the Z-magnification of an atomic force microscope were identified to estimate the uncertainty according to ISO/IEC Guide 98-3:2008. The relative expanded uncertainties ranged from 4.7% to 5.9%.
Source: openaire · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Force Microscopy Techniques and Applications · Metal and Thin Film Mechanics · Surface and Thin Film Phenomena

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References