Improvement in lifetime of pseudo-Schottky diode sensor: Towards selective detection of O3 in a gaseous mixture (O3, NO2)
Source abstract
Abstract
Abstract This article concerns a pseudo-Schottky diode Palladium/Indium-Phosphide (Pd–InP) gas sensor. The catalytic activity of such a palladium layer coupled with a pseudo-Schottky structure enables the measurement of very low concentrations of two highly oxidant gases: nitrogen dioxide (NO 2 ) and ozone (O 3 ). The submission of the sensor to long O 3 exposures leads to a degradation of its sensor characteristics (response time, recovery time and sensitivity) due to oxidation of the palladium metallization by O 3 . Therefore, to improve sensor lifetime and reduce drift, a methodology based on cyclic regeneration of the sensor's palladium surface (carbon monoxide (CO) reduction associated with thermal treatment) has been developed. The pseudo-Schottky gas sensor associated with this methodology exhibits reproducible responses, significant resolution and real time detection in the range of 20–100 ppb for NO 2 and O 3 . Moreover, a sensor exposed to 20 ppb of O 3 presented twice the response of the same sensor exposed to 100 ppb of NO 2 (10.5 nA for 6.5 nA). Selectivity towards O 3 , with this methodology, is demonstrated in the case of atmospheric pollution monitoring.
Source: openaire · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Gas Sensing Nanomaterials and Sensors · Analytical Chemistry and Sensors · Advanced Chemical Sensor Technologies
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References
Reference metadata was not supplied for this record.