Potentiometric cholesterol biosensor based on ZnO nanorods chemically grown on Ag wire
Source abstract
Abstract
An electrochemical biosensor based on ZnO nanorods for potentiometric cholesterol determination is proposed. Hexagon-shaped ZnO nanorods were directly grown on a silver wire having a diameter of 250 mu m using low temperature aqueous chemical approach that produced ZnO nanorods with a diameter of 125250 nm and a length of similar to 1 mu m. Cholesterol oxidase (ChOx) was immobilized by a physical adsorption method onto ZnO nanorods. The electrochemical response of the ChOx/ZnO/Ag biosensor against a standard reference electrode (Ag/AgCl) was investigated as a logarithmic function of the cholesterol concentration (1 x 10(-6)M to 1 x 10(-2)M) showing good linearity with a sensitivity of 35.2 mV per decade and the stable output signal was attained at around 10 s. (C) 2010 Elsevier BM. All rights reserved. (Less)
Source: openaire · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Electrochemical sensors and biosensors · Electrochemical Analysis and Applications · Analytical Chemistry and Sensors
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References