Validation of vital sign measurement by nanosecond pulse near-field sensing and the conductive silver paste printed textile technique
Source abstract
Abstract
Vital sign monitoring is fundamental in daily patient care. Wearable technology has become an important tool in personalized medicine; however, there is a tradeoff problem between accuracy and comfort caused by the elastic textile used. Our study sought to examine the accuracy of heart rate (HR) and breathing rate (BR) monitoring as measured by iSmartweaR, a novel wearable and wireless sensor that uses the laser Doppler technique, with special focus on sex and different static postures. A total of 30 male and 23 female patients, including 23 patients with cardiovascular diseases and 30 healthy participants, were recruited. The HR and BR were measured with iSmartweaR and the gold standard method (CardioSoft and MasterScreen™ CPX) simultaneously in three static postures (supine rest, right lateral decubitus, and left lateral decubitus). The mean errors of HR and BR between iSmartweaR and the gold standard method were close to zero (0.55 and 0.71, respectively). The standard deviations of the HR error and BR error were 0.59 and 0.67, respectively. Disease, posture, sex, and measurement time did not cause clinically significant measurement errors. The overall percentage of missing data was 7.01%. The use of iSmartweaR was valid across different sexes and static postures. It is real-time, continuous and manpower-saving, and can be applied in patient care.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Non-Invasive Vital Sign Monitoring · Heart Rate Variability and Autonomic Control · Hemodynamic Monitoring and Therapy
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Department of Physical Medicine & Rehabilitation, Chang Gung Memorial Hospital, Linkou, 333, Taiwan
Industrial Technology Research Institute, Hsinchu, 310401, Taiwan
Eclat Textile Co., Ltd.,Taiwan, New Taipei City 248, Taiwan
Department of Physical Medicine and Rehabilitation, New Taipei Municipal Tucheng Hospital, Chang Gung Memorial Hospital, New Taipei City, 236, Taiwan
College of Medicine, Chang Gung University, Kwei-Shan, Tao-Yuan County, 333, Taiwan
Healthy Aging Research Center, Chang Gung University, Taiwan
References (36)
Show source-provided references
1. 10.1186/s13049-016-0213-810.1186/s13049-016-0213-8 ↗
2. 10.1111/ijn.1232910.1111/ijn.12329 ↗
3. 10.1081/CEH-20003195910.1081/CEH-200031959 ↗
4. 10.1186/cc813210.1186/cc8132 ↗
5. 10.1007/s00421-007-0552-210.1007/s00421-007-0552-2 ↗
6. 10.1001/archinte.161.22.270910.1001/archinte.161.22.2709 ↗
7. Yuan G · McMaster Univ Med J · 2013 · 10 · 23
8. 10.1007/s10877-014-9575-510.1007/s10877-014-9575-5 ↗
9. Leuvan CH · Crit Care Resusc · 2008 · 10 · 111
10. 10.1016/j.hrthm.2017.03.03810.1016/j.hrthm.2017.03.038 ↗
11. Lin HD · Google Patents · 2019
12. 10.1126/science.125016910.1126/science.1250169 ↗
13. Bouwstra S · Annu Int Conf IEEE Eng Med Biol Soc · 2011 · 2011 · 2488
14. Ciani O · Annu Int Conf IEEE Eng Med Biol Soc · 2008 · 2008 · 1292
15. Wang Z · Sensors (Basel) · 2017
16. 10.3390/rs1315290510.3390/rs13152905 ↗
17. 10.3390/s1707162210.3390/s17071622 ↗
18. Jähne-Raden N, et al. Heartbeat the odds: a novel digital ballistocardiographic sensor system. In: Rasit Eskicioglu (Ed.) proceedings of the 15th ACM conference on embedded network sensor systems , Paper number 77, Delft, Netherlands, November 6-8, 2017, pp. 1–2. New York: US.10.1145/3131672.3136973 ↗
19. 10.3390/s1911249210.3390/s19112492 ↗
20. Klap T · Comput Cardiol · 2013
21. Ali E, Suratman FY. Ultra wideband radar for respiratory monitoring on sleep position. J Meas Electron Commun Syst 2021; 8: 28–34.
22. 10.3390/s2010299910.3390/s20102999 ↗
23. 10.1145/326495810.1145/3264958 ↗
24. 10.3390/bios604005410.3390/bios6040054 ↗
25. 10.1016/S0300-9572(02)00098-910.1016/S0300-9572(02)00098-9 ↗
26. 10.1136/bmj.324.7334.38710.1136/bmj.324.7334.387 ↗
27. 10.1016/j.maturitas.2019.03.01210.1016/j.maturitas.2019.03.012 ↗
28. 10.1016/S0140-6736(20)30925-910.1016/S0140-6736(20)30925-9 ↗
29. 10.1136/bmjhci-2019-10007910.1136/bmjhci-2019-100079 ↗
30. ANSI/AAMI EC13:2002. Cardiac monitors, heart rate meters, and alarms.
31. Food U · 2007
32. 10.3390/jpm702000310.3390/jpm7020003 ↗
33. 10.3109/03091902.2010.50330810.3109/03091902.2010.503308 ↗
34. 10.1136/bmjopen-2017-02016210.1136/bmjopen-2017-020162 ↗
35. 10.1007/s00380-019-01347-810.1007/s00380-019-01347-8 ↗
36. 10.3390/s2004101310.3390/s20041013 ↗