Tomography based on detection of noncollimated backscattered radiation after single-sided photon injection
Source abstract
Abstract
The possibility of using single-sided photon injection to reconstruct a tested object structure on the basis of the intensity of backscattered noncollimated radiation is considered. The reconstruction algorithm is based on statistical simulation of the interaction between radiation and a homogenous object, whose geometry is adequate for that of the tested object. The results of reconstruction are discussed.
Source: openalex_inverted_index · Quality: baseline
Research context
OpenAlex topics: Advanced X-ray and CT Imaging · Medical Imaging Techniques and Applications · Photoacoustic and Ultrasonic Imaging
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (8)
Show source-provided references
1. Gorshkov, V.A., A Tomograph on Noncollimiated Scattered Radiation, Defektoskopiya, 1999, no. 9, pp. 22–28.
2. Gorshkov, V.A., Kröning, M., Anosov, Yu.V., and Dzhogrochoo, O., Tomografiya na rasseyannom nekollimirovannom izluchenii (Tomography on Scattered Noncollimated Radiation), Moscow: Tekhnopoligraftsentr, 2002.
3. Gorshkov, V.A., Kröning, M., Anosov, Y.V., Kuzelev, N.R., Yumashev, V.M., Samossadny, V.T., and Miloserdin, V.Yu., X-ray Tomography on the Base of Noncollimated Radiation, Computer Tomography and Image Processing for Industrial Radiology, Berlin: 2003, p. 9.
4. Roye, W., Niemann, W., and Fischer, K.-H., The X-ray Backscattering Tomography ComScan, Int. Symposium on Computerized Tomography for Industrial Application, Berlin: 1994, pp. 136–140.
5. Gorshkov, V.A., Kuzelev, N.R., Anosov, Yu.V., Yumashev, V.M., Miloserdin, V.Yu., and Samosadnyi, V.T., Detection of Infantry-Protection Bombs Using a Tomograph on Backscattering Noncollimated Irradiation, 3-ya mezhdunarodnaya konferentsiya “Nerazrushayushchii kontrol’ i tekhnicheskaya diagnostika v promyshlennosty” (3d Int. Conf. on Nondestructive Testing and Technical Diagnostics in Industry), Moscow, 2004, pp. 169–170.
6. Leipunskii, O.I., Novozhilov, B.V., and Sakharov, V.N., Propagation of Gamma Quanta in Matter, Fiz. Met. Metalloved., 1960, p. 207.
7. Gorshkov, V.A., Method of Tomographic Evaluation of Density Distribution and Effective Atomic Number in Matter, RF Patent no. 2168717, 2001, p. 5.
8. Gorshkov, V.A., Tomographic Reconstruction of Distributions of Effective Atomic Numbers of Light Elements, Defektoskopiya, 2001, no. 9, pp. 73–80 [Rus. J. of Nondestructive Testing (Engl. Transl.), 2001, vol. 37, no. 9, p. 660].10.1023/A:1014947608746 ↗