Simulation Model for Evaluation of Nonmonoenergeticity and Scattering Artifacts in Computer Tomography
Source abstract
Abstract
A simulation model is proposed for evaluating nonmonoergeticity and scattering artifacts in computed tomography in parallel beam geometry using the example of inhomogeneous objects with axial symmetry. The modeling algorithm consists of a block for generating projections and a block for reconstructing images of sections of an object based on the inverse Abel transform. The developed algorithm is implemented as a MathCad program. The efficiency of the algorithm and software is demonstrated by the example of a multilayer sphere. The closeness of the model and experimental estimates of the radial distributions of the linear radiation attenuation coefficient is proved.
Source: openalex_inverted_index · Quality: baseline
Research context
OpenAlex topics: Advanced X-ray and CT Imaging · Medical Imaging Techniques and Applications · Radiation Dose and Imaging
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (55)
Show source-provided references
1. Kak, A.C., Computerized tomography with X-ray, emission, and ultrasound sources, Proc. IEEE, 1979, vol. 67, no. 9, pp. 1245–1272. https://doi.org/10.1109/PROC.1979.1144010.1109/PROC.1979.11440 ↗
2. Hsieh, J., Computed Tomography: Principles, Design, Artifacts, and Recent Advances, Bellingham: SPIE, 2009.
3. Sugawara, H., Takayanagi, T., Ishikawa, T., Katada, Y., Fukui, R., Yamamoto, Y., and Suzuki, S., New fast kVp switching dual-energy CT: Reduced severity of beam hardening artifacts and improved image quality in reducediodine virtual monochromatic imaging, Acad. Radiol., 2020, vol. 27, no. 11, pp. 1586–1593. https://doi.org/10.1016/j.acra.2019.11.01510.1016/j.acra.2019.11.015 ↗
4. Hur, J., Kim, D., Shin, Y.G., and Lee, H., Metal artifact reduction method based on a constrained beam-hardening estimator for polychromatic x-ray CT, Phys. Med. Biol., 2021, vol. 66, article ID 66.065025. https://doi.org/10.1088/1361-6560/abe02610.1088/1361-6560/abe026 ↗
5. Stolfi, A., De Chiffre, L., and Kasperl, S., Error sources, in Industrial X-Ray Computed Tomography, Cham: Springer, 2018, pp. 143–184. https://doi.org/10.1007/978-3-319-59573-3_510.1007/978-3-319-59573-3_5 ↗
6. Barrett, F. and Keat, N., Artifacts in CT: Recognition and avoidance, Radiographics, 2004, vol. 24, no. 6, pp. 1679–1691. https://doi.org/10.1148/rg.24604506510.1148/rg.246045065 ↗
7. Alvarez, R.E. and Macovski, A., Energy-selective reconstructions in x-ray computerised tomography, Phys. Med. Biol., 1976, vol. 21, no. 5, pp. 733–744. https://doi.org/10.1088/0031-9155/21/5/00210.1088/0031-9155/21/5/002 ↗
8. Kuchenbecker, S., Faby, S., Sawal, S., Lell, M., and Kachelrieß, M., Dual energy CT: How well can pseudomonochromatic imaging reduce metal artifacts?, Med. Phys., 2015, vol. 42, no. 2, pp. 1023– 1036. https://doi.org/10.1118/1.490510610.1118/1.4905106 ↗
9. Li, L., Chen, Z., Wang, G., Chu, J., and Gao, H., A tensor PRISM algorithm for multi-energy CT reconstruction and comparative studies, J. X-ray Sci. Technol., 2014, vol. 22, no. 2, pp. 147–163. https://doi.org/10.3233/xst-14041610.3233/xst-140416 ↗
10. Jacobsen, M.C., Duan, X., Cody, D.D., Cressman, E., Schellingerhout, D., and Layman, R.R., Determination of the limit of detection for iodinated contrast agents with multi-energy computed tomography, in Medical Imaging 2018: Physics of Medical Imaging, Houston, 2018, vol. 10573, article ID 105734Q. https://doi.org/10.1117/12.229393510.1117/12.2293935 ↗
11. Si-Mohamed, S., Bar-Ness, D., Sigovan, M., Cormode, D.P., Coulon, P., Coche, E., Vlassenbroek, A., Normand, G., Boussel, L., and Douek, P., Review of an initial experience with an experimental spectral photoncounting computed tomography system, Nucl. Instrum. Methods Phys. Res., Sect. A, 2017, vol. 873, pp. 27–35. https://doi.org/10.1016/j.nima.2017.04.01410.1016/j.nima.2017.04.014 ↗
12. Raja, A.Y., Moghiseh, M., Bateman, C.J., De Ruiter, N., Schon, B., Schleich, N., Woodfield, T.B.F., Butler Anthony, P.H., and Anderson, N.G., Measuring identification and quantification errors in spectral CT material decomposition, Appl. Sci., 2018, vol. 8, no. 3, article ID 467. https://doi.org/10.3390/app803046710.3390/app8030467 ↗
13. Abella, M., Martínez, C., Desco, M., Vaquero, J.J., and Fessler, J.A., Simplified statistical image reconstruction for X-ray CT with beam-hardening artifact compensation, IEEE Trans. Med. Imaging, 2019, vol. 39, no. 1, pp. 111–118. https://doi.org/10.1109/TMI.2019.292192910.1109/TMI.2019.2921929 ↗
14. Tang, S., Huang, K., Cheng, Y., Mou, X., and Tang, X., Optimization based beam-hardening correction in CT under data integral invariant constraint, Phys. Med. Bio., 2018, vol. 63, no. 13, article ID 135015. https://doi.org/10.1088/1361-6560/aaca1410.1088/1361-6560/aaca14 ↗
15. Runje, B., Keran, Z., and Orošnjak, M., Image artefacts in industrial computed tomography, Tehn. Glasnik, 2020, vol. 14, no. 4, pp. 434–439. https://doi.org/10.31803/tg-2020042712231310.31803/tg-20200427122313 ↗
16. Zbijewski, W. and Beekman, F.J., Efficient Monte Carlo based scatter artifact reduction in cone-beam micro-CT, IEEE Trans. Med. Imaging, 2006, vol. 25, no. 7, pp. 817–827. https://doi.org/10.1109/TMI.2006.87232810.1109/TMI.2006.872328 ↗
17. Sabo-Napadensky, I. and Amir, O., Reduction of scattering artifact in multislice CT, in Medical Imaging 2005: Physics of Medical Imaging, San Diego, 2005, vol. 5745, pp. 983–992. https://doi.org/10.1117/12.59488510.1117/12.594885 ↗
18. Nikolaev, D., Buzmakov, A., Chukalina, M., Yakimchuk, I., Gladkov, A., and Ingacheva, A., CT image quality assessment based on morphometric analysis of artifacts, 2016 Int. Conf. Robotics Mach. Vision (Moscow, 2017), vol. 10253, article ID 102530B. https://doi.org/10.1117/12.226626810.1117/12.2266268 ↗
19. Zav’yalkin, F.M. and Osipov, S.P., Effects of background radiation on image reconstruction error in computerized tomography, Sov. At. Energ., 1991, vol. 70, no. 3, pp. 228–231. https://doi.org/10.1007/BF0112647210.1007/BF01126472 ↗
20. Park, H.S., Choi, J.K., and Seo, J.K., Characterization of metal artifacts in X-ray computed tomography, Commun. Pure Appl. Math., 2017, vol. 70, no. 11, pp. 2191–2217. https://doi.org/10.1002/cpa.2168010.1002/cpa.21680 ↗
21. Yue, D., Fan Rong, C., Ning, C., Liang, H., Ai Lian, L., Ru Xin, W., and Ya Hong, L., Reduction of metal artifacts from unilateral hip arthroplasty on dual-energy CT with metal artifact reduction software, Acta Radiol., 2018, vol. 59, no. 7, pp. 853–860. https://doi.org/10.1177/028418511773147510.1177/0284185117731475 ↗
22. Hsieh, J., Gurmen, O.E., and King, K.F., Investigation of a solid-state detector for advanced computed tomography, IEEE Trans. Med. Imaging, 2000, vol. 19, no. 9, pp. 930–940. https://doi.org/10.1109/42.88784010.1109/42.887840 ↗
23. Zefreh, K.Z., Welford, F.M., and Sijbers, J., Investigation on the effect of exposure time on scintillator afterglow for ultra-fast tomography acquisition, J. Instrum., 2016, vol. 11, no. 12, article ID C12014. https://doi.org/10.1088/1748-0221/11/12/C1201410.1088/1748-0221/11/12/C12014 ↗
24. Miller, S.R., Bhandari, H.B., Bhattacharya, P., Brecher, C., Crespi, J., Couture, A., Dinca, C., Rommel, M., Nagarkar, V.V., Reduced afterglow codoped CsI: Tl for high-energy imaging, IEEE Trans. Nucl. Sci., 2018, vol. 65, no. 8, pp. 2105–2108. https://doi.org/10.1109/TNS.2018.280798610.1109/TNS.2018.2807986 ↗
25. Brönnimann, C. and Trüb, P., Hybrid pixel photon counting X-ray detectors for synchrotron radiation, in Synchrotron Light Sources and Free-Electron Lasers: Accelerator Physics, Instrumentation and Science Applications, Cham: Springer, 2020, pp. 1191–1223. https://doi.org/10.1007/978-3-319-14394-1_3610.1007/978-3-319-14394-1_36 ↗
26. Cammin, J., Kappler, S.G., Weidinger, T., and Taguchi, K., Evaluation of models of spectral distortions in photon-counting detectors for computed tomography, J. Med. Imaging, 2016, vol. 3, no. 2, article ID 023503. https://doi.org/10.1117/1.JMI.3.2.02350310.1117/1.JMI.3.2.023503 ↗
27. Danielsson, M., Persson, M., and Sjölin, M., Photon-counting x-ray detectors for CT, Phys. Med. Biol., 2021, vol. 66, no. 3, article ID 03TR01. https://doi.org/10.1088/1361-6560/abc5a510.1088/1361-6560/abc5a5 ↗
28. Rathee, S.D., Koles, Z.J., and Overton, T.R., Image restoration in computed tomography: Estimation of the spatially variant point spread function, IEEE Trans. Med. Imaging, 1992, vol. 11, no. 4, pp. 539–545. https://doi.org/10.1109/42.19268910.1109/42.192689 ↗
29. Wang, W., Gang, G.J., Siewerdsen, J.H., and Stayman, J.W., Spatial resolution and noise prediction in flatpanel cone-beam CT penalized-likelihood reconstruction, in Medical Imaging 2018: Physics of Medical Imaging, Houston, 2018, vol. 10573, article ID 1057346. https://doi.org/10.1117/12.229454610.1117/12.2294546 ↗
30. Xu, Y., Bai, T., Yan, H., Ouyang, L., Pompos, A., Wang, J., Zhou, L., Jiang, S.B., and Jia, X., A practical conebeam CT scatter correction method with optimized Monte Carlo simulations for image-guided radiation therapy, Phys. Med. Biol., 2015, vol. 60, no. 9, article ID 3567. https://doi.org/10.1088/0031-9155/60/9/356710.1088/0031-9155/60/9/3567 ↗
31. Busi, M., Olsen, U.L., Knudsen, E.B., Frisvad, J.R., Kehres, J., Dreier, E.S., Khalil, M., and Haldrup, K., Simulation tools for scattering corrections in spectrally resolved x-ray computed tomography using McXtrace, Opt. Eng., 2018, vol. 57, no. 3, article ID 037105. https://doi.org/10.1117/1.OE.57.3.03710510.1117/1.OE.57.3.037105 ↗
32. Thomsen, M., Knudsen, E.B., Willendrup, P.K., Bech, M., Willner, M., Pfeiffer, F., Poulsen, M., Lefmann, K., and Feidenhans, R., Prediction of beam hardening artefacts in computed tomography using Monte Carlo simulations, Nucl. Instrum. Methods Phys. Res., Sect. B, 2015, vol. 342, pp. 314–320. https://doi.org/10.1016/j.nimb.2014.10.01510.1016/j.nimb.2014.10.015 ↗
33. Peterzol, A., Létang, J.M., and Babot, D., A beam stop based correction procedure for high spatial frequency scatter in industrial cone-beam X-ray CT, Nucl. Instrum. Methods Phys. Res., Sect. B, 2008, vol. 266, no. 18, pp. 4042–4054. https://doi.org/10.1016/j.nimb.2008.07.00510.1016/j.nimb.2008.07.005 ↗
34. Nakano, T., and Nakashima, Y., Analytical expressions for the reconstructed image of a homogeneous cylindrical sample exhibiting a beam hardening artifact in X-ray computed tomography, J. X-ray Sci. Technol., 2018, vol. 26, no. 5, pp. 691–705. https://dx.doi.org/10.3233%2FXST-18378
35. Osipov, S.P., Yadrenkin, I.G., Chakhlov, S.V., Osipov, O.S., Usachev, E.Yu., and Manushkin, A.A., Calcula ion model of X-ray computed tomography with density assessment function, Russ. J. Nondestr. Test., 2021, vol. 57, no. 3, pp. 222–237. https://doi.org/10.1134/S106183092103004910.1134/S1061830921030049 ↗
36. Kramers, H.A., XCIII, The X-ray spectrum absorption, Philos. Mag., 1923, vol. 46, no. 275, pp. 836–871. https://doi.org/10.1080/1478644230856524410.1080/14786442308565244 ↗
37. Zink, F.E., X-ray tubes, Radiographics, 1997, vol. 17, no. 5, pp. 1259–1268. https://doi.org/10.1148/radiographics.17.5.930811310.1148/radiographics.17.5.9308113 ↗
38. Sasaya, T., Sunaguchi, N., Hyodo, K., Zeniya, T., and Yuasa, T., Multi-pinhole fluorescent x-ray computed tomography for molecular imaging, Sci. Rep., 2017, vol. 7, no. 1, pp. 1–12. https://doi.org/10.1038/s41598-017-05179-210.1038/s41598-017-05179-2 ↗
39. Sidulenko, O.A., Kas’yanov, V.A., Kas’yanov, S.V., and Osipov, S.P., Estimated efficiency of slit collimation of a high-energy radiation source for radiometric testing of large objects, Russ. J. Nondestr. Test., 2006, vol. 42, no. 2, pp. 101–105. https://doi.org/10.1134/S106183090602005710.1134/S1061830906020057 ↗
40. Benetskii, B.A. and Plotnikova, M.V., Gamma-radiation accumulation factors for composite materials and radiation shields, Bull. Lebedev Phys. Inst., 2012, vol. 39, no. 4, pp. 113–117. https://doi.org/10.3103/S106833561204004510.3103/S1068335612040045 ↗
41. Osipov, S.P., Prischepa, I.A., Chakhlov, S.V., Osipov, O.S., and Usachev, E.Yu., Algorithms for modeling the formation and processing of information in X-ray tomography of foam materials, Russ. J. Nondestr. Test., 2021, vol. 57, no. 3, pp. 238–250. https://doi.org/10.1134/S106183092103005010.1134/S1061830921030050 ↗
42. Zhang, T., Xing, Y., Zhang, L., Jin, X., Gao, H., and Chen, Z., Stationary computed tomography with source and detector in linear symmetric geometry: Direct filtered backprojection reconstruction, Med. Phys., 2020, vol. 47, no. 5, pp. 2222–2236. https://doi.org/10.1002/mp.1405810.1002/mp.14058 ↗
43. Miqueles, E., Koshev, N., and Helou, E.S., A backprojection slice theorem for tomographic reconstruction, IEEE Trans. Image Proc., 2017, vol. 27, no. 2, pp. 894–906. https://doi.org/10.1109/TIP.2017.276678510.1109/TIP.2017.2766785 ↗
44. Brooks, R.A. and Di Chiro, G., Theory of image reconstruction in computed tomography, Radiology, 1975, vol. 117, no. 3, pp. 561–572. https://doi.org/10.1148/117.3.56110.1148/117.3.561 ↗
45. Nishimura, M., Psaltis, D., Caimi, F., and Casasent, D., Implementation of the inverse Radon transform by optical convolution, Opt. Commun., 1978, vol. 25, no. 3, pp. 301–304. https://doi.org/10.1016/0030-4018(78)90133-510.1016/0030-4018(78)90133-5 ↗
46. Gustafsson, B., Mathematics for computer tomography, Phys. Scr., 1996, vol. 1996, no. T61, pp. 38–43. https://doi.org/10.1088/0031-8949/1996/T61/00610.1088/0031-8949/1996/T61/006 ↗
47. Schofield, R., King, L., Tayal, U., Castellano, I., Stirrup, J., Pontana, F., and Nicol, E., Image reconstruction: Part 1—understanding filtered back projection, noise and image acquisition, J. Cardiovasc. Comp. Tomogr., 2020, vol. 14, no. 3, pp. 219–225. https://doi.org/10.1016/j.jcct.2019.04.00810.1016/j.jcct.2019.04.008 ↗
48. Dolmatova, A., Chukalina, M., and Nikolaev, D., Accelerated FBP for computed tomography image reconstruction, 2020 IEEE Int. Conf. Image Process. (ICIP) (Anchorage, 2020), pp. 3030–3034. https://doi.org/10.1109/ICIP40778.2020.919104410.1109/ICIP40778.2020.9191044 ↗
49. Shepp, L.A. and Logan, B.F., The Fourier reconstruction of a head section, IEEE Trans. Nucl. Sci., 1974, vol. 21, no. 3, pp. 21–43. https://doi.org/10.1109/TNS.1974.649923510.1109/TNS.1974.6499235 ↗
50. Osipov, S., Chakhlov, S., Batranin, A., Osipov, O., and Kytmanov, J., Theoretical study of a simplified implementation model of a dual-energy technique for computed tomography, NDT & E Int., 2018, vol. 98, pp. 63–69. https://doi.org/10.1016/j.ndteint.2018.04.01010.1016/j.ndteint.2018.04.010 ↗
51. Ametova, E., Ferrucci, M., Chilingaryan, S., and Dewulf, W., A computationally inexpensive model for estimating dimensional measurement uncertainty due to x-ray computed tomography instrument misalignments, Meas. Sci. Technol., 2018, vol. 29, no. 6, article ID 065007. https://doi.org/10.1088/1361-6501/aab1a110.1088/1361-6501/aab1a1 ↗
52. De Micheli, E., A fast algorithm for the inversion of Abel’s transform, Appl. Math. Comp., 2017, vol. 301, pp. 12–24. https://doi.org/10.1016/j.amc.2016.12.00910.1016/j.amc.2016.12.009 ↗
53. Gholami, M., Rashedi, A., Lenoir, N., Hautemayou, D., Ovarlez, G., and Hormozi, S., Time-resolved 2D concentration maps in flowing suspensions using X-ray, J. Rheol., 2018, vol. 62, no. 4, pp. 955–974. https://doi.org/10.1122/1.499406310.1122/1.4994063 ↗
54. De Hoop, M.V. and Ilmavirta, J., Abel transforms with low regularity with applications to x-ray tomography on spherically symmetric manifolds, Inverse Probl., 2017, vol. 33, no. 12, article ID 124003. https://doi.org/10.1088/1361-6420/aa942310.1088/1361-6420/aa9423 ↗
55. Harima, Y., An approximation of gamma-ray buildup factors by modified geometrical progression, Nucl. Sci. Eng., 1983, vol. 83, no. 2, pp. 299–309. https://doi.org/10.13182/NSE83-A1822210.13182/NSE83-A18222 ↗