F-14 Waste Reduction Process Improvements in the Analysis of Plutonium by X-ray Fluorescence: Results from Multiple Data Sets
Source abstract
Abstract
To minimize waste, improve process safety, and minimize costs, modifications were implemented to a method for quantifying gallium in plutonium metal using wavelength dispersive X-ray fluorescence. These changes included reducing sample sizes, reducing ion exchange process volumes, using cheaper reagent grade acids, eliminating the use of HF acid, and using more robust containment film for sample analysis. Relative precision and accuracy achieved from analyzing multiple aliquots from a single parent sample were {approx}0.2% and {approx}0.1% respectively. The same precision was obtained from analyzing a total of four parent materials, and the average relative accuracy from all the samples was 0.4%, which is within programmatic uncertainty requirements.
Source: openalex_inverted_index · Quality: baseline
Research context
OpenAlex topics: Radioactive element chemistry and processing · Radioactive contamination and transfer · Nuclear Physics and Applications
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References
Reference metadata was not supplied for this record.