Cover to provide an inert atmosphere for XRD sample changers
Source abstract
Abstract
A cover for Scintag's six and twelve-position sample changers was designed and constructed to provide an inert atmosphere for samples during diffraction batch runs. The cover is equipped with inlet and outlet gas ports and fits over the top of the sample changer. Using dry nitrogen gas fed into the inlet of the cover, a sample of lithium bromide was protected from atmospheric moisture for greater than 18 h. The cover uses a thin Mylar window that gives greater than 95% transparency for copper K -alpha X-rays. The cover is a simple device that allows our lab to run multiple moisture-sensitive samples in a batch mode. The simple approach and materials used in the construction of the cover could be applied to other brands of powder diffractometers.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: X-ray Diffraction in Crystallography · Nuclear Physics and Applications · Advanced X-ray Imaging Techniques
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (2)
Show source-provided references
1. Saturated salt solutions for maintaining specified relative humidities
2. 10.1017/S088571560001072110.1017/S0885715600010721 ↗