journal article·1999-12

Environmentally controlled sample stage for Scintag's six-position XRD sample changer

Charles Buckner, Christopher Kidd, Stacey Long, Peter Varlashkin, Jim Zhu

Powder Diffraction· Volume 14· Issue 4
Publisher page

Abstract

An environmentally controlled sample stage for Scintag's six-position XRD sample changer is described. Unlike the commercial environmental stages for Scintag XRD units which require removal of the sample changer, and installation and alignment of the sample stage, the stage described herein is designed to fit directly on the sample changer with no realignment. Such a stage provides for more convenient operation for an X-ray diffractometer with multiple users in both routine (ambient environment) and nonroutine (nonambient temperatures up to 250 °C and varying relative humidities) modes. In addition, the new stage uses a window made of a material that is safer to machine compared to beryllium (which is toxic).
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Nuclear Physics and Applications · X-ray Diffraction in Crystallography · Advanced X-ray and CT Imaging

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (3)

Show source-provided references
1. Byrn, S. R. (1982). Solid State Chemistry of Drugs (Academic, New York), pp. 3–27.
2. 10.1107/S002188989301382210.1107/S0021889893013822 ↗
3. 10.1016/0378-5173(96)04601-710.1016/0378-5173(96)04601-7 ↗