Improving energy stability in the National Institute of Standards and Technology Microcalorimeter X-ray detector
Source abstract
Abstract
Microcalorimeter X-ray detectors using a transition edge sensor depend for their linearity and energy scale on the stability of the operating point on the transition curve. We report on some sources of energy scale drift in microcalorimeter X-ray detectors and the manner in which they have been addressed. Previously observed drifts of >10 eV∕h have been reduced to 1–2 eV∕h. This improved stability has resulted in the observation of X-ray fluorescence linewidths of ⩽12 eV over 6 h of counting time.
Source: openalex_inverted_index · Quality: baseline
Research context
OpenAlex topics: Superconducting and THz Device Technology · Physics of Superconductivity and Magnetism · Microwave Engineering and Waveguides
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (3)
Show source-provided references
1. 10.1063/1.36715310.1063/1.367153 ↗
2. 10.1046/j.1365-2818.1997.2670824.x10.1046/j.1365-2818.1997.2670824.x ↗
3. 10.1016/0375-9474(91)90238-210.1016/0375-9474(91)90238-2 ↗