Quantitative analysis with the transition edge sensor microcalorimeter X-ray detector

Terrence Jach, Nicholas Ritchie, Joel Ullom, James A. Beall

Powder Diffraction· Volume 22· Issue 2
Publisher page

Abstract

We report on the use of a microcalorimeter X-ray detector with a transition edge sensor in an electron probe to perform quantitative analysis. We analyzed two bulk samples of multielement glasses that have been previously characterized by chemical methods for use as standard reference materials. The spectra were analyzed against standards using three different correction schemes. In one of the standards, the reference line was easily resolved despite its proximity within 45 eV of another line. With the exception of direct measurements of oxygen (a particularly challenging element), the results are in agreement with the certified characterization to better than 1% absolute or 8% relative. This demonstrates the potential of microcalorimeter detectors as replacements for conventional energy dispersive detectors in applications requiring high energy resolution.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Superconducting and THz Device Technology · Physics of Superconductivity and Magnetism · Surface and Thin Film Phenomena

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (15)

Show source-provided references
1. 10.1016/S0168-9002(99)01354-610.1016/S0168-9002(99)01354-6 ↗
2. 10.1046/j.1365-2818.1997.2670824.x10.1046/j.1365-2818.1997.2670824.x ↗
3. 10.1103/RevModPhys.75.3510.1103/RevModPhys.75.35 ↗
4. Pouchou J. L. , Pichoir F. , and Boivin D. (1989). “Further improvements in quantitation procedures for X-ray microanalysis,” The 12th International Congress on X-ray Optics and Microanalysis, Cracow, Poland, August–September 1989.10.1002/sca.4950120407 ↗
5. 10.1007/s00604-003-0131-010.1007/s00604-003-0131-0 ↗
6. 10.1016/0375-9474(91)90238-210.1016/0375-9474(91)90238-2 ↗
7. 10.1002/(SICI)1097-4539(199801/02)27:1<3::AID-XRS227>3.0.CO;2-L10.1002/(SICI)1097-4539(199801/02)27:1<3::AID-XRS227>3.0.CO;2-L ↗
8. 10.1016/j.nima.2005.12.14610.1016/j.nima.2005.12.146 ↗
9. 10.1063/1.36715310.1063/1.367153 ↗
10. 10.1016/j.nima.2005.12.12110.1016/j.nima.2005.12.121 ↗
11. Improving energy stability in the National Institute of Standards and Technology microcalorimeter X-ray detector
12. 10.1002/xrs.51210.1002/xrs.512 ↗
13. 10.1002/sia.209310.1002/sia.2093 ↗
14. NIST (1981). Certificate of Analysis, Standard Reference Material 470, Mineral Glasses for Microanalysis (National Institute of Standards and Technology, Gaithersburg, MD), 〈http://ts.nist.gov/MeasurementServices/ReferenceMaterials/ARCHIVED_CERTIFICATES/470.pdf〉.
15. NIST (1984). Certificate of Analysis, Standard Reference Material 1873 (National Institute of Standards and Technology, Gaithersburg, MD).