Towards Improved Alignment of Powder Diffractometers
Source abstract
Abstract
Abstract An improved method of alignment of a Philips diffractometer equipped with a theta-compensating slit is described. The method employs a special fluorescent screen which is flat to better than 10 microns. The procedure results in accurate alignment of the theta-compensating slit at low angles and assures reliable intensity data down to one degree two theta or less.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: X-ray Diffraction in Crystallography
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (9)