FD‐7 Dosimeters Irradiated in Different Radiation Environments: A Comparison Between Monte Carlo Simulations and Experiments
Source abstract
Abstract
This work presents a comparison between experimental data and Monte Carlo simulations realized with particle and heavy ion transport code system to evaluate radiation transport in commercial FD‐7 radiophotoluminescent (RPL) glass dosimeters exposed to different radiation environments. The analysis is performed for photon and electron, representative of experimental irradiation conditions encountered in research and industrial facilities where RPL dosimeters are employed. Photon sources include Co‐60 gamma rays, X‐ray tubes with tungsten targets operating at 100, 160, and 225 kV, and high‐energy photons produced by 6 MeV electrons impinging on a tungsten target. Monoenergetic electron beams of 2.5 and 6 MeV are also studied. For each configuration, the depth‐dose distributions within the RPL dosimeter are analyzed to evaluate dose homogeneity across the detector volume. Conversion factors between absorbed dose in RPL and reference materials commonly used in dosimetry, such as water, air, and silica, are calculated. Experimental data from various irradiation campaigns realized using the mentioned sources are presented and used to validate the simulation methodology. Agreement between simulated and measured doses is found, supporting the validity of the modeling approach. The results provide a tool to better design irradiation test campaigns and for the interpretation of RPL readouts in a variety of irradiation conditions.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Radiation Detection and Scintillator Technologies · Radiation Therapy and Dosimetry · Advanced Radiotherapy Techniques
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Institut d’Optique Graduate School Laboratoire Hubert Curien UMR 5516 CNRS Université Jean Monnet Saint‐Etienne France
European Organization for Nuclear Research (CERN) Geneva Switzerland
CEA DAM CEG Gramat France
Laboratoire des Solides Irradiés CEA CNRS École polytechnique Institut Polytechnique de Paris Palaiseau France
Department of Physics and Chemistry Emilio Segrè University of Palermo Palermo Italy
Institut Universitaire de France (IUF) Ministère de l’Enseignement Supérieur et de la Recherche Paris France
References (48)
Show source-provided references
1. 10.1063/1.357616910.1063/1.3576169 ↗
2. К.Becker RADIOPHOTOLUMINESCENT DOSIMETRY.
3. 10.1269/jrr.41.12910.1269/jrr.41.129 ↗
4. 10.1118/1.186115510.1118/1.1861155 ↗
5. Becker K. · 1973 · 141
6. 10.1093/rpd/ncp19510.1093/rpd/ncp195 ↗
7. Huang D. Y. · 2011
8. 10.7567/JJAP.57.10640210.7567/JJAP.57.106402 ↗
9. 10.35848/1347-4065/ac902310.35848/1347-4065/ac9023 ↗
10. 10.1016/j.radmeas.2020.10636310.1016/j.radmeas.2020.106363 ↗
11. 10.1109/TNS.2022.317478410.1109/TNS.2022.3174784 ↗
12. 10.1016/j.optmat.2010.04.00410.1016/j.optmat.2010.04.004 ↗
13. 10.1016/j.radmeas.2019.02.00910.1016/j.radmeas.2019.02.009 ↗
14. 10.1093/rpd/ncm15710.1093/rpd/ncm157 ↗
15. 10.1016/j.radmeas.2020.10626310.1016/j.radmeas.2020.106263 ↗
16. 10.1103/PhysRevAccelBeams.25.10300110.1103/PhysRevAccelBeams.25.103001 ↗
17. 10.1109/TNS.2024.336527210.1109/TNS.2024.3365272 ↗
18. 10.1016/j.radmeas.2024.10724610.1016/j.radmeas.2024.107246 ↗
19. 10.1109/TNS.2025.354994910.1109/TNS.2025.3549949 ↗
20. 10.1109/TNS.2025.354998710.1109/TNS.2025.3549987 ↗
21. 10.1016/j.radmeas.2020.10624610.1016/j.radmeas.2020.106246 ↗
22. 10.1016/j.radmeas.2020.10627810.1016/j.radmeas.2020.106278 ↗
23. Chiyoda Technol Corporation Dose ACE RPL In‐Vivo Dosimetry System for Medical Use Technical Data Sheet Dec. 2021 accessed September 03 2023.https://www.c‐technol.co.jp/en/.
24. 10.1016/j.radmeas.2024.10720410.1016/j.radmeas.2024.107204 ↗
25. A.Nourreddine Study of a New Neutron Dosimeter Incorporating RPL Detectors.
26. 10.1016/j.radmeas.2007.11.01210.1016/j.radmeas.2007.11.012 ↗
27. 10.1109/TNS.2025.354645910.1109/TNS.2025.3546459 ↗
28. S.Girard et al. ‘PETRA: an Experimental Platform for X‐ray Radiation Testing of Materials and Components’ submitted to SiO2 conference’ (2025).
29. 10.1109/TNS.2023.323995010.1109/TNS.2023.3239950 ↗
30. J.Lefèvre O.Cavani andB.Boizot “The SIRIUS facility: A powerful tool for studying irradiation effects in materials ” in13th International Topical Meeting of Nuclear Applications of Accelerators(2017).
31. 10.1109/RADECS.2016.809315810.1109/RADECS.2016.8093158 ↗
32. M.Cecchetto et al. Dosimetry in a Mixed Spallation Source with RPLs and FLUKA Simulations.
33. 10.1109/TNS.2018.282986410.1109/TNS.2018.2829864 ↗
34. 10.1080/00223131.2023.227573610.1080/00223131.2023.2275736 ↗
35. 10.1002/mp.1494510.1002/mp.14945 ↗
36. Attix F. H. · 2004
37. Giovanni Santin ESA / ESTEC and RheaTech Ltd Geant4 tutorial Paris 4‐8 June 2007. Accessed: Feb. 25 2025 http://geant4.in2p3.fr/2007/prog/GiovanniSantin/GSantin_Geant4_Paris07_Normalisation_v07.pdf.
38. Giovanni Santin ESA / ESTEC and RheaTech Ltd On behalf of the Geant4 collaboration Ecole Geant4 Annecy 18‐21 and 25‐28 Nov 2008 accessed February 25 2025 https://www.fe.infn.it/u/paterno/Geant4_tutorial/slides_further/GPS/GSantin_Geant4_Annecy2008_GPS_v11.pdf.
39. 10.3390/radiation301000610.3390/radiation3010006 ↗
40. A. K.Alem Dosimetry Study in an X‐Ray Irradiator: Monte Carlo Simulations and Experimental Results on Radio‐Photoluminescence Samples(Master's thesis EMJMD RADMEP 2023) https://jyx.jyu.fi/handle/123456789/89326.
41. IAEA Consultancy Meeting “Radiation effects on polymer materials ” inIAEA Int. Atomic Energy Agency Vienna Austria Meeting Rep. EVT1905348 (2019) http://www‐naweb.iaea.org/.
42. 10.1016/j.radphyschem.2004.03.06210.1016/j.radphyschem.2004.03.062 ↗
43. Barlow R. J. · 1993
44. NIST X‐ray’. [Online] https://www.nist.gov/pml/x‐ray‐mass‐attenuation‐coefficients.
45. NIST electrons [Online] https://www.nist.gov/pml/stopping‐power‐range‐tables‐electrons‐protons‐and‐helium‐ions.
46. 10.1109/TNS.2024.351145510.1109/TNS.2024.3511455 ↗
47. 10.1021/acsami.4c2122010.1021/acsami.4c21220 ↗
48. 10.1088/2515‐7639/addb7d10.1088/2515‐7639/addb7d ↗