Performance of high-TC dc SQUID magnetometers for use in a magnetically disturbed environment
Source abstract
Abstract
YBCO dc SQUID magnetometers based on bicrystal Josephson junctions on 10 mm × 10 mm STO substrates have been fabricated. We have designed three different types of pickup coils for the SQUID magnetometers, i.e., solid-type, 12 and 16 parallel loops with 50 μm line width to test performances of the SQUIDs for use in a magnetically disturbed environment. Magnetometer with 16-parallel-loop pickup coil exhibit most stable FLL operation under external dc magnetic field. Finally, we could obtain optimised direct coupled YBCO SQUID magnetometer design having flux transfer coefficient BΦ of 4.5 nT/Φ0 and magnetic field noise BN of 30 fT/Hz1/2 measured at 100 Hz. (© 2004 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim)
Source: openalex_inverted_index · Quality: baseline
Research context
OpenAlex topics: Physics of Superconductivity and Magnetism · Atomic and Subatomic Physics Research · Magnetic Field Sensors Techniques
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References
Reference metadata was not supplied for this record.