YBCO dc SQUID magnetometers based on bicrystal Josephson junctions on 10 mm × 10 mm STO substrates have been fabricated. We have designed three different types of pickup coils for the SQUID magnetometers, i.e., solid-type, 12 and 16 parallel loops with 50 μm line width to test performances of the SQUIDs for use in a magnetically disturbed environment. Magnetometer with 16-parallel-loop pickup coil exhibit most stable FLL operation under external dc magnetic field. Finally, we could obtain optimised direct coupled YBCO SQUID magnetometer design having flux transfer coefficient BΦ of 4.5 nT/Φ0 and magnetic field noise BN of 30 fT/Hz1/2 measured at 100 Hz. (© 2004 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim)