A multichannel double relaxation oscillation SQUID magnetometer system for magnetoencephalogram
Source abstract
Abstract
We developed multichannel SQUID magnetometer and gradiometer systems based on double relaxation oscillation SQUIDs (DROS) for use in biomagnetic studies, such as magnetoencephalogram (MEG). DROS consists of two SQUIDs (signal SQUID and reference SQUID) in series, and a relaxation circuit of an inductor and a resistor. Specially, we used single reference junction instead of the reference SQUID. Since DROS magnetometer has about 10 times lager flux-to-voltage transfer coefficient than dc SQUID, we can use simple flux-locked loop electronics for SQUID operation. The measured noise at 100 Hz was 3 fT/ for the magnetometer and 4 fT/ for the planar gradiometer, which are low enough for neuromagnetic measurements. We also constructed two kinds of multichannel SQUID systems, 40-channel and 37-channel system. Using this system, we measured several MEG signals. (© 2004 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim)
Source: openalex_inverted_index · Quality: baseline
Research context
OpenAlex topics: Atomic and Subatomic Physics Research · Functional Brain Connectivity Studies · Advanced MRI Techniques and Applications
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References
Reference metadata was not supplied for this record.