An analytic expression is derived which determines the electric field strength E(r) in the sensitive region of an one-open-face coaxial detector (spheric coaxial detector - SC detector). The cases of no charge in the sensitive region (Na-Nd = 0) and of uniform charge distribution (Na-Nd = = const. ± 0) are considered. The applicability of the expression obtained is confirmed experi-mentally for a model SC detector. The E(r) dependence in SC detectors is shown to be steeper than in coaxial detectors with two open faces (right coaxial detector - RC detector). This de-pendence is found to steepen when approaching the closed face and near it and to be steepest along the cylinder axis.