Laser induced desorption studied with mirror electron microscopy

R. Gauthier, M Babout, C Guittard

Publisher page

Abstract

The study of the laser induced desorption suggests to associate this technics to the mirror electron microscopy for measuring the work function changes and the total electron reflection coefficient at a metallic surface, during the sample desorption. The results are concerned with W (100) single crystal samples. A pulsed ruby laser light is used with an energy in the range 0.01 to 3 J. The measurements are performed with a mirror electron microscope, using very slow electrons (0 to 5 eV) in the vicinity of the surface sample. The apparatus is inside an ultrahigh vacuum chamber (10−9 to 10−10 Torr); a quadrupole mass spectrometer (QMS) may either be used as a detector of laser produced ions or as a residual gas analyzer; a microprocessor associated to a lock-in amplifier stores I(U) and dI(U)/dU (where I and U are respectively the backscattered electron current and the bias voltage of the sample); the use of a deconvolution treatment eliminates the effect of the electron spread of the incident electrons. So, the reflection coefficient can be measured after laser irradiation: the values observed show the efficiency of the laser desorption.