article·1997-04-01Open access

Influence of Sample Pretreatment Methods on the Determination of Trace Oxygen in Iron and Steel

Hisao Yasuhara, Makoto Shimura, Keiichi Yoshioka

Abstract

Removing the oxide film from the sample surface is an important process in analyzing trace oxygen in iron and steel. The depth profiles of oxygen on the surface after electropolishing in CH3COOH— HClO4, CrO3—H3PO4, and HCl, and chemical polishing by HF– H2O2 are investigated. The thickness of the oxide layer after electropolishing is estimated to be about 3 nm, which is approximately equivalent to 1 mass ppm oxygen. This indicates that the currently available surface polishing methods are inadequate for the analysis of mass ppm or sub-ppm oxygen levels. A surface without an oxide layer is obtained by Ar ion sputtering in a vacuum, but the oxide layer reappears within 1 min when the sample comes into contact with air.
Source: openalex_inverted_index · Quality: baseline

Research context

OpenAlex topics: Analytical Chemistry and Sensors · Water Quality Monitoring and Analysis · Advanced Chemical Sensor Technologies

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.