article·1970-06-16

A modification of the theory of the triple crystal X-ray spectrometer and its application for practical calculations

K. Godwod

Publisher page

Abstract

The paper describes a certain modification of the theory of the triple crystal X-ray spectrometer. The most general case, the (nI, nII, nIII) arrangement of the spectrometer, is discussed. The interference functions of each of the three crystals are described in the coordinates in which the dependence on the angular position of the crystals is given by the dynamical theory of X-ray diffraction. The final formula enables us to carry out direct numerical calculations of the shapes of rocking curves for the triple crystal X-ray spectrometer. Die Arbeit enthält die Entwicklung einer Drei-Kristall-Röntgenspektrometer-Theorie. Ein allgemeiner Fall der Spektrometeranordnung (nI, nII, nIII) wird diskutiert. Die Interferenzfunktion jedes Kristalls wird in den Koordinaten, in denen die Abhängigkeit der Intensitätsverbreitung von der Winkelposition der Kristalle durch die dynamische Theorie der Röntgstrahlbeugung gegeben wird, beschrieben. Die Endformel erlaubt es, die numerischen Berechnungen der Form der „rocking-Kurven”︁ für Drei-Kristall-Röntgenspektrometer direkt durchzuführen.
Source: openalex_inverted_index · Quality: baseline

Research context

OpenAlex topics: X-ray Diffraction in Crystallography · X-ray Spectroscopy and Fluorescence Analysis

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References