journal article·2026-10Open access

Characterization of Ce:La-GPS scintillator for tritiated water detection

Ichiro Nishii, Tomone Suganuma, Takahiko Horiai, Yuui Yokota, Masao Yoshino, Makoto Nakajima, Masanori Hara, Keisuke Shigemori, Kohei Yamanoi

Optical Materials· Volume 178

Abstract

Tritium in water is difficult to monitor because the low-energy beta particles emitted by tritium are strongly self-absorbed. Although liquid scintillation counting provides high sensitivity, it requires sample preparation and generates radioactive organic liquid waste, limiting its suitability for on-site continuous monitoring. In this study, we investigated the inorganic single-crystal scintillator (La,Gd) 2 Si 2 O 7 :Ce (Ce:La-GPS) for tritiated water measurements. The as-grown sample showed long-term afterglow that increased the count rate in the region of interest for tritium detection, but air annealing at 1200 °C suppressed the afterglow to the background level. The annealed sample exhibited a light output of approximately 35,000 photons/MeV and an energy resolution of 5.5% at 662 keV under 137 Cs gamma-ray excitation. Immersion measurements in tritiated water showed a linear counting response, with a minimum detectable activity of 3.85 MBq/L for a 600 s measurement. These results indicate that Ce:La-GPS is a promising solid scintillator for high-activity tritiated water monitoring, while application to low-level monitoring will require increasing the effective scintillator surface area.
Source: openalex_inverted_index · Quality: baseline

Research context

OpenAlex topics: Radiation Detection and Scintillator Technologies · Radioactive contamination and transfer · Astro and Planetary Science

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (25)

Show source-provided references
1. Tritium production, recovery and application in Korea10.1016/j.apradiso.2009.02.078 ↗
2. Development of plant concept related to tritium handling in the water-cooling system for JA DEMO10.1016/j.fusengdes.2019.03.174 ↗
3. Guidelines for drinking-water quality
4. ESTAR: stopping-Power and range tables for electrons
5. Knoll · 2010 · 359
6. Surette · 1986
7. Detector for tritium in water10.1016/0029-554X(70)90596-3 ↗
8. An on-line tritium-in-water monitor10.1016/0168-9002(85)90141-X ↗
9. Development of a practical tritiated water monitor to supervise the discharge of treated water from Fukushima Daiichi nuclear power plant10.1016/j.nima.2024.169208 ↗
10. Knoll · 2010 · 226
11. A plastic scintillator-based activity monitor for tritiated water in the GBq/L range10.1016/j.nima.2015.06.040 ↗
12. Comparative testing of various flow-cell detectors fabricated using CaF2 solid scintillator10.13182/FST14-T39 ↗
13. Development of continuous measurement method for tritium in water based on CaF2(Eu) particles and wavelength shift fibers
14. The continuous measurement system of tritium for total amount of ALPS treated water at Fukushima nuclear power plant, Adv. X-Ray
15. Knoll · 2010 · 245
16. Composition engineering in cerium-doped (Lu,Gd)3(Ga,Al)5O12 single-crystal scintillators10.1021/cg200694a ↗
17. Evaluation of GAGG:Ce scintillators for future space applications10.1088/1748-0221/13/02/P02023 ↗
18. State of the art challenges and prospects of advanced materials in radiation detection for nuclear energy: a review10.1016/j.pmatsci.2025.101616 ↗
19. Fast and high-energy-resolution oxide scintillator: Ce-doped (La,Gd)2Si2O7
20. Scintillation properties of Ce:(La,Gd)2Si2O7 at high temperatures10.1016/j.nima.2014.10.008 ↗
21. Growth, structural considerations, and characterization of Ce-doped (La,Gd)2Si2O7 scintillating crystals10.1021/cg501416u ↗
22. Czochralski growth of 2 in. Ce-doped (La,Gd)2Si2O7 for scintillator application10.1016/j.jcrysgro.2016.06.016 ↗
23. The divalent ion codoping effect on Ce-doped (Gd, La)2Si2O7 single crystals10.1016/j.optmat.2017.02.062 ↗
24. Crystal structure of Ce-doped (La,Gd)2Si2O7 grown by the Czochralski process10.1016/j.jallcom.2018.02.286 ↗
25. Limits for qualitative detection and quantitative determination application to radiochemistry10.1021/ac60259a007 ↗