journal article·2013-05-10Open access

Calibration of oscillation amplitude in dynamic scanning force microscopy

Juan Francisco González Martínez, Inés Nieto-Carvajal, Jaime Colchero

Nanotechnology· Volume 24· Issue 18

Abstract

A method to precisely calibrate the oscillation amplitude in dynamic scanning force microscopy is described. It is shown that the typical electronics used to process the dynamic motion of the cantilever can be adjusted to transfer the thermal noise of the cantilever motion from its resonance frequency to a much lower frequency within the typical bandwidth of the corresponding data acquisition electronics of a scanning force microscopy system. Based on this concept, two procedures for the calibration of the oscillation amplitude are proposed. One is based on a simple calculation of the root mean square deviation measured at the outputs of the electronics used to process the dynamic motion of the cantilever, and the second one is based on analysis of the corresponding spectrum and the calculation of the quality factor, the resonance frequency and the signal strength.We show that the proposed scheme for amplitude calibration using thermal noise is experimentally and theoretically robust, with soft as well as with hard cantilevers. Moreover, it is directly related to well-defined quantities such as the force constant and thermal energy, in contrast to the calibration using amplitude versus distance curves, which requires non-trivial a priori assumptions regarding the amplitude versus distance relation.
Source: openalex_inverted_index · Quality: baseline

Research context

OpenAlex topics: Force Microscopy Techniques and Applications · Piezoelectric Actuators and Control · Scientific Measurement and Uncertainty Evaluation

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (29)

Show source-provided references
1. 10.1063/1.34734710.1063/1.347347 ↗
2. 10.1063/1.35234810.1063/1.352348 ↗
3. 10.1103/RevModPhys.75.94910.1103/RevModPhys.75.949 ↗
4. 10.1016/S0167-5729(02)00077-810.1016/S0167-5729(02)00077-8 ↗
5. 10.1063/1.47942210.1063/1.479422 ↗
6. Simon G H · Nanotechnology · 2007 · 1810.1088/0957-4484/18/25/255503 ↗
7. 10.1063/1.12143410.1063/1.121434 ↗
8. 10.1103/PhysRevLett.82.338810.1103/PhysRevLett.82.3388 ↗
9. 10.1016/j.susc.2012.05.00910.1016/j.susc.2012.05.009 ↗
10. 10.1103/PhysRevLett.96.03610710.1103/PhysRevLett.96.036107 ↗
11. 10.1016/S1369-7021(10)70162-010.1016/S1369-7021(10)70162-0 ↗
12. Jamitzky F · Nanotechnology · 2006 · 17 · S21310.1088/0957-4484/17/7/S19 ↗
13. 10.1103/PhysRevB.66.04140610.1103/PhysRevB.66.041406 ↗
14. 10.1186/1556-276X-7-17410.1186/1556-276X-7-174 ↗
15. 10.6028/jres.116.01510.6028/jres.116.015 ↗
16. 10.1016/j.ultramic.2011.09.01210.1016/j.ultramic.2011.09.012 ↗
17. Burnham N A · Nanotechnology · 2003 · 14 · 110.1088/0957-4484/14/1/301 ↗
18. 10.1063/1.10102410.1063/1.101024 ↗
19. Colchero J · Procedures in Scanning Probe Microscopies · 1998 · 133
20. 10.1063/1.114397010.1063/1.1143970 ↗
21. Butt H J · Nanotechnology · 1995 · 6 · 110.1088/0957-4484/6/1/001 ↗
22. 10.1016/S0304-3991(00)00077-210.1016/S0304-3991(00)00077-2 ↗
23. Malegori G · Nanotechnology · 2011 · 2210.1088/0957-4484/22/19/195702 ↗
24. Malegori G · Atomic Force Microscopy Imaging, Measuring and Manipulating Surfaces at the Atomic Scale · 2012
25. 10.1063/1.36572610.1063/1.365726 ↗
26. 10.1063/1.353376910.1063/1.3533769 ↗
27. Stanford Research · 2010
28. Stanford Research · 2009
29. 10.1063/1.115002110.1063/1.1150021 ↗