journal article·1998-06-01

In situself-calibration of atomic force microscopy

Hyun-Kyu Kweon, Wei Gao, Satoshi Kiyono

Nanotechnology· Volume 9· Issue 2
Publisher page

Abstract

This paper introduces an in situ self-calibration method to correct Z-directional distortion associated with atomic force microscopy (AFM), which occurs mainly due to the nonlinear motion of the servo piezo-electric elements (PZT) of the AFM. The principle of the method, results of simulation and experiment are presented. In this method, the derivative of the calibration curve function of the PZT actuator is calculated from two sets of profile measurement data, which are obtained from two measurements with a small Z-direction shift. Simulation results showed that the Z-directional distortion can be compensated with an accuracy approximately twice that of the Z-directional resolution of the calibrated AFM. The effectiveness of the proposed self-calibration method was confirmed through experimental calibration of a commercially available AFM.
Source: openalex_inverted_index · Quality: baseline

Research context

OpenAlex topics: Force Microscopy Techniques and Applications · Piezoelectric Actuators and Control · Advanced MEMS and NEMS Technologies

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (8)

Show source-provided references
1. 10.1016/0141-6359(95)00002-U10.1016/0141-6359(95)00002-U ↗
2. Okayama S · J. Japan Soc. Precis. Eng. · 1988 · 54 · 81710.2493/jjspe.54.817 ↗
3. 10.1063/1.114250610.1063/1.1142506 ↗
4. Sasaki A · J. Japan Soc. Precis. Eng. · 1993 · 59 · 9510.2493/jjspe.59.95 ↗
5. 10.1063/1.114704610.1063/1.1147046 ↗
6. Kiyono S · 1996 · 186
7. Kiyono S · 1997
8. Gao W · 1997 · 282