Abstract
Abstract Several years ago this laboratory reported on the effects of stray light in an infrared microscope and its consequences on the spatial resolution achievable in a typical line scan or mapping experiment of a polymer film. In this study it was demonstrated that unless the size of the sampled area was greater than twice the wavelength of light employed for the analysis, a significant amount of stray light from near neighbors was contained in the sampled areas. Further, this generalization is only valid for sample sizes above the diffraction limit. The results of this study are summarized in Table. The ratio of the sampled area to that of the aperture size (corresponding to expected sample size) can be thought of as a degradation factor in spatial resolution. In this earlier study the sample geometry was chosen to minimize stray light effects arising from the sample size and shape.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Calibration and Measurement Techniques · Infrared Target Detection Methodologies · Spectroscopy and Chemometric Analyses
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Molecular Microspectroscopy Laboratory, Dept. of Chemistry, Miami University , Oxford, OH 45056
References (2)
Show source-provided references
1. Sommer · Applied Spectroscopy · 1991 · 45 · 163310.1366/0003702914335229 ↗
2. Reffner · 1995