A Shield for Reducing the Thermal Signal from Heating Holders DuringIn SituEnergy-Dispersive X-Ray Spectroscopy Analysis
Source abstract
Abstract
This article describes a simple shield that can be placed on typical commercial heating holders to reduce the thermal signal during heating to reasonable levels forin situenergy-dispersive X-ray spectroscopy analysis. The improved temperature capability provided by the shield is demonstrated by initial compositional analysis results obtained across a solid–liquid interface on Al-Si-Cu-Mg alloy powder particles. Considerations in the design of and improvement for the shield are discussed.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Aluminum Alloy Microstructure Properties · X-ray Diffraction in Crystallography · Advanced ceramic materials synthesis
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (4)
Show source-provided references
1. Storaska, G.A. & Howe, J.M. (2004).In-situ transmission electron microscopy investigation ofsurface-oxide, stress-relief mechanisms during melting of sub-micrometerAl–Si alloy particles.Mater Sci Eng A368,183–190.10.1016/j.msea.2003.10.315 ↗
2. Kamino, T. , Sasaki, K. & Saka, H. (1997).High resolution electron microscopy in situ observation ofdynamic behavior of grain boundaries and interfaces at very hightemperatures.Microsc Microanal 3,393–408.10.1017/S143192769797029X ↗
3. Valimet, Inc. (2004).Home page available at: http://www.valimet.com/.