Characterization of SiGe Films for Use as a National Institute of Standards and Technology Microanalysis Reference Material (RM 8905)
Source abstract
Abstract
Abstract Bulk silicon-germanium (SiGe) alloys and two SiGe thick films (4 and 5 μm) on Si wafers were tested with the electron probe microanalyzer (EPMA) using wavelength dispersive spectrometers (WDS) for heterogeneity and composition for use as reference materials needed by the microelectronics industry. One alloy with a nominal composition of Si0.86Ge0.14and the two thick films with nominal compositions of Si0.90Ge0.10and Si0.75Ge0.25on Si, evaluated for micro- and macroheterogeneity, will make good microanalysis reference materials with an overall expanded heterogeneity uncertainty of 1.1% relative or less for Ge. The bulk Ge composition in the Si0.86Ge0.14alloy was determined to be 30.228% mass fraction Ge with an expanded uncertainty of the mean of 0.195% mass fraction. The thick films were quantified with WDS-EPMA using both the Si0.86Ge0.14alloy and element wafers as reference materials. The Ge concentration was determined to be 22.80% mass fraction with an expanded uncertainty of the mean of 0.12% mass fraction for the Si0.90Ge0.10wafer and 43.66% mass fraction for the Si0.75Ge0.25wafer with an expanded uncertainty of the mean of 0.25% mass fraction. The two thick SiGe films will be issued as National Institute of Standards and Technology Reference Materials (RM 8905).
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Electron and X-Ray Spectroscopy Techniques · Surface and Thin Film Phenomena · Scientific Measurement and Uncertainty Evaluation
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (35)
Show source-provided references
1. Marinenko · 2003 · 238
2. 10.1007/978-1-4899-2617-3_1510.1007/978-1-4899-2617-3_15 ↗
3. McMaster W.H. , Del Grande N.K. , Mallet J.H. & Hubbell J.H. (1969). Compilation of X-ray cross sections. Report UCRL-50174, Lawrence Livermore Laboratory.
4. Silicon-germanium economic drivers, technology, and volume production issues
5. Application of alpha-factor and Monte-Carlo methods to epma in the system Ge-Si10.1017/S1431927600029482 ↗
6. 10.6028/jres.105.04710.6028/jres.105.047 ↗
7. 10.1021/ac015509710.1021/ac0155097 ↗
8. CITZAF: A package of correction programs for the quantitative electron microbeam X-ray analysis of thick polished materials, thin films, and particles
9. 10.1002/sca.495019010110.1002/sca.4950190101 ↗
10. 10.1023/A:100672272905110.1023/A:1006722729051 ↗
11. Chantler · 2005
12. 10.1021/ac053354n10.1021/ac053354n ↗
13. 10.1007/978-1-4613-9975-9_1410.1007/978-1-4613-9975-9_14 ↗
14. 10.1016/0092-640X(82)90002-X10.1016/0092-640X(82)90002-X ↗
15. Heinrich · 1986 · 67
16. Pouchou · 1988 · 315
17. 2003
18. 10.1007/s10967-007-0209-y10.1007/s10967-007-0209-y ↗
19. 10.1002/sia.209310.1002/sia.2093 ↗
20. 10.1086/62733910.1086/627339 ↗
21. 10.1007/978-1-4899-2617-3_410.1007/978-1-4899-2617-3_4 ↗
22. 10.1007/BF0210987910.1007/BF02109879 ↗
23. Berger · 2005
24. Donovan J. (2005). Probe Software, Inc. Web site at http://www.probesoftware.com.
25. Fiori C.E. & Swyt-Thomas C.R. (1991). U.S. Patent number 5,299,138, accepted 1994. Desk top spectrum analyzer. Free version available at http://www.cstl.nist.gov/div837/Division/outputs/software.htm.
26. Heinrich · 1966 · 296
27. 10.1063/1.34272010.1063/1.342720 ↗
28. 1995
29. 10.1017/S143192760404038310.1017/S1431927604040383 ↗
30. NIST micro MC: A user's guide to the NIST microanalysis Monte Carlo electron trajectory simulation program
31. 10.1039/b715148a10.1039/b715148a ↗
32. 10.1021/ac000087710.1021/ac0000877 ↗
33. Scott · 1983
34. 10.1007/978-1-4899-2617-3_310.1007/978-1-4899-2617-3_3 ↗
35. 10.6028/NIST.TN.129710.6028/NIST.TN.1297 ↗