journal article·2003-04

Charge-Related Problems Associated with X-Ray Microanalysis in the Variable Pressure Scanning Electron Microscope at Low Pressures

Brendan J. Griffin, Alexandra A. Suvorova

Microscopy and Microanalysis· Volume 9· Issue 2
Publisher page

Abstract

In variable pressure scanning electron microscopy (VPSEM) the current data suggests that considerable caution is required in the interpretation of X-ray data from nonconductive samples, depending on the operating conditions. This article reviews some of the documented approaches and presents data that illustrate the nature and magnitude of the effects of charge above, on, and in the sample on the detected X-ray emissions from the sample and from elsewhere within the VPSEM specimen chamber. The collection of reliable and reproducible X-ray data has been found to require relatively high specimen chamber gas pressures, at the upper end of or beyond the available pressures for most VPSEMs. It is also shown that sample characteristics, including composition, strongly influence local charge effects, which can significantly affect the primary electron landing energy and consequently the resultant emitted X-ray signal under low pressure environments.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Electron and X-Ray Spectroscopy Techniques · Corrosion Behavior and Inhibition · Advanced Electron Microscopy Techniques and Applications

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References

Reference metadata was not supplied for this record.