An Iterative Qualitative–Quantitative Sequential Analysis Strategy for Electron-Excited X-ray Microanalysis with Energy Dispersive Spectrometry: Finding the Unexpected Needles in the Peak Overlap Haystack

Dale E. Newbury, Nicholas W. M. Ritchie

Microscopy and Microanalysis· Volume 24· Issue 4
Publisher page

Abstract

Abstract When analyzing an unknown by electron-excited energy dispersive X-ray spectrometry, with the entire periodic table possibly in play, how does the analyst discover minor and trace constituents when their peaks are overwhelmed by the intensity of an interfering peak(s) from a major constituent? In this paper, we advocate for and demonstrate an iterative analytical approach, alternating qualitative analysis (peak identification) and standards-based quantitative analysis with peak fitting. This method employs two “tools”: (1) monitoring of the “raw analytical total,” which is the sum of all measured constituents as well as any such as oxygen calculated by the method of assumed stoichiometry, and (2) careful inspection of the “peak fitting residual spectrum” that is constructed as part of the quantitative analysis procedure in the software engine DTSA-II (a pseudo-acronym) from the National Institute of Standards and Technology. Elements newly recognized after each round are incorporated into the next round of quantitative analysis until the limits of detection are reached, as defined by the total spectrum counts.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Electron and X-Ray Spectroscopy Techniques · X-ray Spectroscopy and Fluorescence Analysis · Ion-surface interactions and analysis

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (17)

Show source-provided references
1. 10.1007/978-1-4939-6676-910.1007/978-1-4939-6676-9 ↗
2. 10.1017/S143192760505053110.1017/S1431927605050531 ↗
3. 10.1017/S143192761700026510.1017/S1431927617000265 ↗
4. 10.1017/S143192761200110910.1017/S1431927612001109 ↗
5. Lifshin E (1974) The use of solid state X-ray detectors for obtaining fundamental X-ray data. Tutorial and Proceedings of the Ninth Annual Microbeam Analysis Society, Ottawa, ON, Canada.
6. 10.1007/978-1-4899-2617-3_410.1007/978-1-4899-2617-3_4 ↗
7. 10.1017/S143192761601183110.1017/S1431927616011831 ↗
8. Reed SJB (1975) The shape of the continuous X-ray spectrum and background corrections for energy-dispersive electron microprobe analysis. X-Ray Spectrometry 4, 14–17.10.1002/xrs.1300040105 ↗
9. 10.1017/S143192761601152110.1017/S1431927616011521 ↗
10. 10.1002/sca.2015110.1002/sca.20151 ↗
11. Armstrong J (2011) Low voltage and low overvoltage X-ray nanoanalysis with field emission electron microprobes and SEMs: Problems in quantitation for first-row transition elements. 2011 AGU Fall Meeting, San Francisco, CA.
12. McCarthy · 1981
13. 10.1017/S143192761400545510.1017/S1431927614005455 ↗
14. Castaing R (1951) Application of electron probes to local chemical and crystallographic analysis. PhD Thesis, University of Paris, Paris, France.
15. 10.1017/S143192761300823410.1017/S1431927613008234 ↗
16. 10.1007/s10853-014-8685-210.1007/s10853-014-8685-2 ↗
17. 10.1017/S143192761700178710.1017/S1431927617001787 ↗