journal article·2023-09-29

Fabrication of Specimens for Atom Probe Tomography Using a Combined Gallium and Neon Focused Ion Beam Milling Approach

Frances I Allen, Paul T Blanchard, Russell Lake, David Pappas, Deying Xia, John A Notte, Ruopeng Zhang, Andrew M Minor, Norman A Sanford

Microscopy and Microanalysis· Volume 29· Issue 5
Publisher page

Abstract

Abstract We demonstrate a new focused ion beam sample preparation method for atom probe tomography. The key aspect of the new method is that we use a neon ion beam for the final tip-shaping after conventional annulus milling using gallium ions. This dual-ion approach combines the benefits of the faster milling capability of the higher current gallium ion beam with the chemically inert and higher precision milling capability of the noble gas neon ion beam. Using a titanium–aluminum alloy and a layered aluminum/aluminum-oxide tunnel junction sample as test cases, we show that atom probe tips prepared using the combined gallium and neon ion approach are free from the gallium contamination that typically frustrates composition analysis of these materials due to implantation, diffusion, and embrittlement effects. We propose that by using a focused ion beam from a noble gas species, such as the neon ions demonstrated here, atom probe tomography can be more reliably performed on a larger range of materials than is currently possible using conventional techniques.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Advanced Materials Characterization Techniques · Hydrogen embrittlement and corrosion behaviors in metals · Fusion materials and technologies

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Department of Materials Science and Engineering, UC Berkeley , Berkeley, CA 94720 , USA

National Center for Electron Microscopy, Molecular Foundry, Lawrence Berkeley National Laboratory , Berkeley, CA 94720 , USA

Physical Measurement Laboratory, National Institute of Standards and Technology , Boulder, CO 80305 , USA

Carl Zeiss SMT Inc. , Danvers, MA 01923 , USA

References (37)

Show source-provided references
1. A review of defect engineering, ion implantation, and nanofabrication using the helium ion microscope10.3762/bjnano.12.52 ↗
2. Nature of gallium focused ion beam induced phase transformation in 316L austenitic stainless steel10.1016/j.actamat.2016.08.008 ↗
3. Bunton · 2022
4. A three-dimensional atom probe microscope incorporating a wavelength-tuneable femtosecond-pulsed coherent extreme ultraviolet light source10.1557/adv.2019.296 ↗
5. Field ion emission in an atom probe microscope triggered by femtosecond-pulsed coherent extreme ultraviolet light10.1017/S1431927620000203 ↗
6. Preparation and analysis of atom probe tips by xenon focused ion beam milling10.1017/S1431927616000581 ↗
7. Xenon plasma focussed ion beam preparation of an Al-6XXX alloy sample for atom probe tomography including analysis of an α-Al(Fe,Mn)Si dispersoid10.1016/j.matchar.2021.111194 ↗
8. Interfaces and defect composition at the near-atomic scale through atom probe tomography investigations10.1557/jmr.2018.375 ↗
9. Atom probe tomography10.1038/s43586-021-00047-w ↗
11. Review article: Review of electrohydrodynamical ion sources and their applications to focused ion beam technology10.1116/1.5047150 ↗
12. Mechanisms of nanodot formation under focused ion beam irradiation in compound semiconductors10.1063/1.3530839 ↗
13. An in-situ approach for preparing atom probe tomography specimens by xenon plasma-focussed ion beam10.1016/j.ultramic.2019.04.005 ↗
14. Atom probe tomography on semiconductor devices10.1002/admi.201500713 ↗
16. LBL CXRO Database · 2023
17. New approach for FIB-preparation of atom probe specimens for aluminum alloys10.1371/journal.pone.0231179 ↗
18. The neon gas field ion source—a first characterization of neon nanomachining properties10.1016/j.nima.2010.12.220 ↗
19. A versatile cryo-transfer system, connecting cryogenic focused ion beam sample preparation to atom probe microscopy10.1371/journal.pone.0245555 ↗
20. Analyzing the channel dopant profile in next-generation FinFETs via atom probe tomography10.1016/j.ultramic.2017.12.013 ↗
21. Evaluation of neon focused ion beam milling for TEM sample preparation10.1111/jmi.12416 ↗
22. Evaporation-field differences with deep-UV atom probe tomography10.1017/S1431927621004736 ↗
23. Modern focused-ion-beam-based site-specific specimen preparation for atom probe tomography10.1017/S1431927616012642 ↗
24. Correlative approach for atom probe sample preparation of interfaces using plasma focused ion beam without lift-out10.1017/S1431927621000349 ↗
25. A review on atom probe and correlative microscopy studies of corrosion in nickel-based superalloys10.1557/s43577-022-00366-7 ↗
26. Investigations of field-evaporated end forms in voltage- and laser-pulsed atom probe tomography10.1016/j.ultramic.2008.10.001 ↗
27. Advances in source technology for focused ion beam instruments10.1557/mrs.2014.53 ↗
28. In situ site-specific specimen preparation for atom probe tomography10.1016/j.ultramic.2006.06.008 ↗
29. Effect of gallium focused ion beam milling on preparation of aluminium thin foils10.1111/j.1365-2818.2010.03401.x ↗
30. High-resolution terahertz-driven atom probe tomography10.1126/sciadv.abd7259 ↗
31. Helium ion microscope: A new tool for nanoscale microscopy and metrology10.1116/1.2357967 ↗
32. Direct observation of the thickness distribution of ultra thin AlOx barriers in Al/AlOx/Al Josephson junctions10.1088/0022-3727/48/39/395308 ↗
33. Comparative apex electrostatics of atom probe tomography specimens10.1007/s11664-021-08932-6 ↗
34. Direct imaging of short-range order and its impact on deformation in Ti–6Al10.1126/sciadv.aax2799 ↗
35. Optimizing atom probe analysis with synchronous laser pulsing and voltage pulsing10.1017/S1431927616012666 ↗
36. Comparing Xe+ pFIB and Ga+ FIB for TEM sample preparation of Al alloys: Minimising FIB-induced artefacts10.1111/jmi.12983 ↗
37. SRIM—The stopping and range of ions in matter (2010)10.1016/j.nimb.2010.02.091 ↗