journal article·2012-08

Ad Hoc Auto-Tuning of Aberrations Using High-Resolution STEM Images by Autocorrelation Function

Hidetaka Sawada, Masashi Watanabe, Izuru Chiyo

Microscopy and Microanalysis· Volume 18· Issue 4
Publisher page

Abstract

Abstract A method for measurement of the aberration status from high-resolution dark-field images is developed using scanning transmission electron microscopy (STEM), called the Segmented Image Autocorrelation function Matrix (SIAM). The method employs an autocorrelation function from the segmented area in the defocused STEM images from an aligned crystalline specimen to measure the defocus and twofold astigmatism for the probe-forming system. The values measured using this method can be fed directly back to the instrument by changing the strength of the stigmator and the objective lens of the microscope. It is successfully demonstrated that the feedback system can automatically correct the defocus and twofold astigmatism of the microscope after several iterations using practical STEM images from an actual crystalline specimen.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Advanced Electron Microscopy Techniques and Applications · Integrated Circuits and Semiconductor Failure Analysis · Image Processing Techniques and Applications

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (15)

Show source-provided references
1. 10.1016/j.ultramic.2005.07.00110.1016/j.ultramic.2005.07.001 ↗
2. 10.1017/S143192761100741010.1017/S1431927611007410 ↗
3. 10.1093/jmicro/dfp03310.1093/jmicro/dfp033 ↗
4. 10.1016/j.ultramic.2008.04.09510.1016/j.ultramic.2008.04.095 ↗
5. 10.1016/j.ultramic.2011.01.03410.1016/j.ultramic.2011.01.034 ↗
6. 10.1016/j.ultramic.2010.04.00610.1016/j.ultramic.2010.04.006 ↗
7. 10.1016/0304-3991(86)90005-710.1016/0304-3991(86)90005-7 ↗
8. 10.1016/S0304-3991(99)00013-310.1016/S0304-3991(99)00013-3 ↗
9. 10.1016/S0304-3991(99)00194-110.1016/S0304-3991(99)00194-1 ↗
10. 10.1038/3382310.1038/33823 ↗
11. Design and development of Cs correctors for 300kV TEM and STEM
12. 10.1016/j.ultramic.2005.06.00710.1016/j.ultramic.2005.06.007 ↗
13. 10.1002/jemt.106003010510.1002/jemt.1060030105 ↗
14. Progress in aberration-corrected scanning transmission electron microscopy