Progress Towards Arrays of Microcalorimeter X-Ray Detectors
Source abstract
Abstract
Abstract The high performance of single-pixel microcalorimeter EDS (μ,cal EDS) has been shown to be very useful for a variety of microanalysis cases. The primary advantage of jxcal EDS over conventional EDS is the factor of 25 improvement in energy resolution (∽3 eV in real-time). This level of energy resolution is particularly important for applications such as nanoscale contaminant analysis where it is necessary to resolve peak overlaps at low x-ray energies. Because μcal EDS offers practical solutions to many microanalysis problems, several companies are proceeding with commercialization of single-pixel μal EDS technology. Two drawbacks limiting the application of uxal EDS are its low count rate (∽500 s−1) and small area (∽0.04 mm for a bare single pixel, ∽5 mm2 with a polycapillary optic). We are developing a 32x32 pixel array with a total area of 40 mm2 and with a total count rate between 105 s−1 and 106 s−1.