On-Line Aberration Measurement and Correction in STEM
Source abstract
Abstract
Abstract Aberration correction in electron microscopy is a subject with a 60 year history dating back to the fundamental work of Scherzer. There have been several partial successes, such as Deltrap's spherical aberration (Cs) corrector which nulled Cs over 30 years ago. However, the practical goal of attaining better resolution than the best uncorrected microscope operating at the same voltage remains to be fulfilled. Combining well-known electron-optical principles with stable electronics, versatile computer control, and software able to diagnose and correct aberrations on-line is at last bringing this goal within reach. We are building a quadrupole-octupole Cs corrector with automated aberration diagnosis for a VG HB5 dedicated scanning transmission electron microscope (STEM). A STEM with no spherical aberration will produce a smaller probe size with a given beam current than an uncorrected STEM, and a larger beam current in a given size probe.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Electron and X-Ray Spectroscopy Techniques · Advanced Electron Microscopy Techniques and Applications · Advancements in Photolithography Techniques
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
MP group,Cavendish Laboratory, University of Cambridge , Cambridge CB3 OHE, UK