NiO Test Specimens for Analytical Electron Microscopy: Round-Robin Results
Source abstract
Abstract
Improvements in instrumentation for energy-dispersive X-ray microanalysis (EDX) and electron energy-loss spectroscopy (EELS) have underlined the need for suitable standards for measuring performance. We report the results from several laboratories that were supplied with a test specimen consisting of a thin film of nickel oxide supported on a molybdenum grid. The Ni-Kα/Mo-Kα count ratio was used as an indication of number of stray electrons and/or X-rays in the TEM column; the Ni-Kα peak/background ratio provided a measure of the total background in the EDX spectrum, including bremsstrahlung contributions and the effect of detector electronics. By providing values typical of current instrumentation, the results illustrate how the test specimen can be used to evaluate TEM/EDX systems prior to purchase, during installation, and (periodically) during operation. The NiO films were also used to test EELS acquisition and quantification procedures: measured Ni/O elemental ratios were all within 10% of stoichiometry.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Electron and X-Ray Spectroscopy Techniques · X-ray Spectroscopy and Fluorescence Analysis · Advanced Electron Microscopy Techniques and Applications
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.