journal article·2014-01

A Removable Hybrid Robot System for Long Bone Fracture Reduction

Tianmiao Wang, Changsheng Li, Lei Hu, Peifu Tang, Lihai Zhang, Hailong Du, Sheng Luan, Lifeng Wang, Yiming Tan, Cheng Peng

Publisher page

Abstract

In traditional long bone fracture reduction surgery, there are some drawbacks such as low accuracy, high radiation for surgeons and a risk of infection. To overcome these disadvantages, a removable hybrid robot system is developed, which integrates a removable series-parallel mechanism with a motor-double cylinder (MDC) driven mode. This paper describes the mechanism in detail, analyses the principle and the method of the fracture reduction, presents the surgical procedure, and verifies the reduction accuracy by experiments with bone models. The results are shown as follows. The mean deviations of the axial displacement and lateral displacement are 1.60mm and 1.26mm respectively. The standard deviations are 0.69mm and 0.30mm. The mean deviations of the side angle and turn inward are 2.06° and 2.22° respectively. The standard deviations are 0.50° and 0.99°. This minimally invasive robot features high accuracy and zero radiation for surgeons, and is able to conduct fracture reduction for long bones.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Soft Robotics and Applications · Prosthetics and Rehabilitation Robotics · Bone fractures and treatments

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

School of Mechanical Engineering and Automation, Beihang University, Beijing, China

Department of Orthopaedics, Chinese PLA General Hospital, Beijing, China

School of Computer Science and Engineering, Beihang University, Beijing, China

References