journal article·2019-05-23

Breakage of metal hip arthroplasty components: Retrieval and structural analysis

Łukasz Łapaj, Waldemar Woźniak, Tomasz Wiśniewski, Justyna Rozwalka, Łukasz Paczesny, Jan Zabrzyński, Piotr Janusz, Jacek Kruczyński

Publisher page

Abstract

BACKGROUND: Breakage of joint arthroplasty components are rare, yet during an implant retrieval program we found several cases. OBJECTIVE: In this study we examined the components to determine the causes and mechanisms of breakage of these implants. METHODS: From our collection of 849 retrievals we selected 682 cases with metal parts (503 hip, 79 knee arthroplasties) and identified fractured components: seven hip resurfacing implants, five total hip replacement stems, one monopolar femoral head, and one modular revision femoral stem from. Implants were examined using optical and scanning electron microscopy; metallographic sections were prepared and samples of periprosthetic tissues underwent microscopic examination. RESULTS: In the resurfacing components breakage occurred in small stems placed in the femoral neck due to necrosis of femoral heads, with no metal flaws detected. Fatigue breakage of femoral stems was caused by presence of material flaws in the CoCrMo alloy, and corrosion. The monopolar head failed in fatigue fracture mechanism, breakage was initiated in an undercut near the taper connection for femoral component. The modular stem from Ti alloy sustained fatigue fracture induced by corrosion caused by debris from previously revised stem; no material flaws were detected in this sample. In most cases periprosthetic tissues had a morphology typical for aseptic loosening. CONCLUSIONS: In our series failure was caused by material flaws, presence of stress raisers and localized corrosion. Our findings indicate that sharp edges and other features which can act as stress raisers should be avoided in newly designed implants. Corrosion induced fracture of the modular Ti stem indicates the need for a detailed debridement of periprosthetic tissues during revision arthroplasties.
Source: openalex_inverted_index · Quality: baseline

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Total Knee Arthroplasty Outcomes · Orthopedic Infections and Treatments

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

, Poznan University of Medical Sciences, , Poland

Metal Forming Institute, , Poland

, , , Poland

References (29)

Show source-provided references
1. Why do some titanium-alloy total hip arthroplasty modular necks fail?10.1016/j.jmbbm.2016.12.012 ↗
2. An analysis of reported cases of fracture of the universal exeter femoral stem prosthesis10.1016/j.arth.2016.09.032 ↗
3. Fractures of modern high nitrogen stainless steel cemented stems: Cause, mechanism, and avoidance in 14 cases10.1016/j.arth.2006.12.051 ↗
4. Femoral component neck fracture after failed hip resurfacing arthroplasty10.1016/j.arth.2010.11.017 ↗
5. New type of hip arthroplasty failure related to modular femoral components: Breakage at the neck-stem junction10.1016/j.otsr.2013.02.010 ↗
6. Fracture of the cobalt-chromium modular femoral neck component in total hip arthroplasty10.1016/j.otsr.2014.03.027 ↗
7. Failure of the modular neck in a total hip arthroplasty10.1016/j.arth.2009.07.015 ↗
8. Pseudotumor from modular neck corrosion after ceramic-on-polyethylene total hip arthroplasty
9. Adverse local tissue reaction arising from corrosion at the femoral neck-body junction in a dual-taper stem with a cobalt-chromium modular neck10.2106/JBJS.L.01042 ↗
10. Early fracture of the modular neck of a MODULUS femoral stem10.1016/j.artd.2016.09.004 ↗
11. Total hip arthroplasty modular neck failure10.1016/j.arth.2010.03.011 ↗
12. Fracture of the tibial component in total knee arthroplasty: Report on two cases
13. Femoral neck fractures following Birmingham hip resurfacing10.1302/0301-620X.87B4.15498 ↗
14. Fracture of cementless femoral stems at the mid-stem junction in modular revision hip arthroplasty systems10.2106/JBJS.I.01589 ↗
15. Corrosion-induced fracture of a double-modular hip prosthesis: A case report10.2106/JBJS.I.00980 ↗
16. Femoral stem fracture and in vivo corrosion of retrieved modular femoral hips10.1016/j.arth.2011.11.007 ↗
17. Removal of the well-bonded distal part of a non-cylindrical broken femoral stem (Autophor 900S) with hollow trephine reamers-report of two cases10.1080/00016470216321 ↗
18. Prosthetic component failures in hip arthroplasty surgery10.1016/S0883-5403(05)80199-8 ↗
19. Revised histopathological consensus classification of joint implant related pathology10.1016/j.prp.2014.09.017 ↗
20. The correlation of wear with histological features after failed hip resurfacing arthroplasty10.2106/JBJS.L.00775 ↗
21. Primary total-hip arthroplasty with the Autophor 900-S fully porous coated stem in young patients: Seven to seventeen years of follow-up10.1016/j.arth.2004.09.048 ↗
22. Fatigue fracture in dual modular revision total hip arthroplasty stems: Failure analysis and computed tomography diagnostics in two cases10.1016/j.arth.2013.09.008 ↗
23. Neck fracture of femoral stems with a sharp slot at the neck: Biomechanical analysis10.1007/s00776-015-0745-1 ↗
24. Corrosion and adverse local tissue reaction in one type of modular neck stem10.1016/j.arth.2015.04.039 ↗
25. Corrosion at the neck-stem junction as a cause of metal ion release and pseudotumour formation. The Journal of bone and joint surgery British volume10.1302/0301-620X.94B7.29122 ↗
26. Failed metal-on-metal hip arthroplasties: A spectrum of clinical presentations and operative findings10.1007/s11999-010-1419-0 ↗
27. Biomechanical, morphological, and histological analysis of early failures in hip resurfacing arthroplasty10.1243/095441105X69015 ↗
28. Osteonecrosis in retrieved femoral heads after failed resurfacing arthroplasty of the hip10.1302/0301-620X.87B3.15330 ↗
29. Corrosion behaviour of polished and sandblasted titanium alloys in PBS solution