journal article·2023-01-11

Analysis of shear stress related hemolysis in a ventricular assist device

Mohamed Bounouib, Hind Benakrach, Mourad Taha-Janan, Wajih Maazouzi

Publisher page

Abstract

BACKGROUND: Implantable devices such as ventricular assist devices provide appropriate treatment for patients with advanced heart failure. Unfortunately, these devices still have many problems, particularly related to blood damage. OBJECTIVE: The aim of this research is to examine two new ventricular assist devices in terms of induced shear stress, exposure time, and induced hemolysis. METHOD: Reverse engineering was used on multiple axial flow ventricular assist devices to collect all the details related to the designs (diameters, lengths, blade angles…), which were used to build two prototypes: Model A and Model B. RESULTS: The obtained results were close to a large extent, except for static pressure rise, where the difference was clear. CONCLUSION: Compared with what has been published in other studies, the overall performance of both models was excellent.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Mechanical Circulatory Support Devices · Cardiac Structural Anomalies and Repair · Blood properties and coagulation

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

, , Mohammed V University, , Morocco

References (23)

Show source-provided references
1. A two-stage rotary blood pump design with potentially lower blood trauma: A computational study10.5301/ijao.5000482 ↗
2. Design and simulation of axial flow maglev blood pump10.5815/ijieeb.2011.02.06 ↗
3. Numerical simulation of an axial blood pump10.1111/j.1525-1594.2007.00422.x ↗
4. Design of implantable axial-flow blood pump and numerical studies on its performance10.1016/S1001-6058(08)60170-5 ↗
5. Design and hemocompatibility analysis of a double-suction injection suspension blood pump using computational fluid dynamics methods: Double-suction injection suspension blood pump10.1111/aor.12888 ↗
6. Blood pump design variations and their influence on hydraulic performance and indicators of hemocompatibility10.1007/s10439-017-1951-0 ↗
7. Design optimization of an axial blood pump with computational fluid dynamics10.1097/MAT.0b013e318164137f ↗
8. A new design for a compact centrifugal blood pump with a magnetically levitated rotor10.1097/01.MAT.0000144364.62671.5A ↗
9. Low flow alarm in a patient with left ventricular assist device: What went wrong?10.1016/S0735-1097(19)32800-1 ↗
10. Device thrombosis in HeartMate II continuous-flow left ventricular assist devices: A multifactorial phenomenon10.1016/j.healun.2013.10.005 ↗
11. Early thrombus in a HeartMate II left ventricular assist device: A potential cause of hemolysis and diagnostic dilemma10.1016/j.jtcvs.2009.09.046 ↗
12. Red blood cell damage by shear stress10.1016/S0006-3495(72)86085-5 ↗
13. Numerical study of a new ventricular assist device10.1111/aor.13635 ↗
14. Three-dimensional numerical prediction of stress loading of blood particles in a centrifugal pump10.1111/j.1525-1594.1995.tb02386.x ↗
15. Estimation of shear stress-related blood damage in heart valve prostheses – in vitro comparison of 25 aortic valves10.1177/039139889001300507 ↗
16. Computational fluid dynamics prediction of blood damage in a centrifugal pump10.1046/j.1525-1594.2003.00026.x ↗
17. Assessment of hemolysis in a ventricular assist axial flow blood pump10.1007/s10527-016-9627-x ↗
18. A Couette viscometer for short time shearing of blood10.3233/BIR-1980-171-205 ↗
19. Study of flow-induced hemolysis using novel Couette-type blood-shearing devices: Thoughts and progress10.1111/j.1525-1594.2011.01243.x ↗
20. Evaluation of shear-induced platelet activation models under constant and dynamic shear stress loading conditions relevant to devices10.1007/s10439-013-0758-x ↗
21. Quantification of shear-induced platelet activation: High shear stresses for short exposure time: Shear-induced platelet activation10.1111/aor.12438 ↗
22. A quantitative comparison of mechanical blood damage parameters in rotary ventricular assist devices: Shear stress, exposure time and hemolysis index10.1115/1.4007092 ↗
23. Numerical analysis of blood damage potential of the HeartMate II and HeartWare HVAD rotary blood pumps: Blood damage potential of blood pumps10.1111/aor.12542 ↗