journal article·1992-11-01

Submonolayer adsorbate reference material based on a low alloy steel fracture sample for Auger electron spectroscopy Part 2 Interlaboratory tests

M. P. Seah, C. P. Hunt, C. M. Younes, G. C. Allen, H. Viefhaus, B. J. Lee, R. K. Wild, H. E. Bishop, T. English

Materials Science and Technology· Volume 8· Issue 11
Publisher page

Abstract

A batch of low alloy reference material for the study of grain boundary segregations of P and Sn has been studied using eight Auger electron spectrometers in six laboratories. The instruments include both cylindrical mirror and concentric hemispherical analysers. Measurements in the differential mode using 5 eV Savitzky and Golay differentiation show that peaks close in energy, such as the ratio of the Sn 430 eV and Fe 651 e V peak-to-peak heights, can be compared directly to demonstrate a reproducibility of 8%. For peaks more widely separated, the scatter of intensity may exceed a factor of 2. However, using reference spectra from the material, all the spectrometers can be calibrated for their intensity/energy response function. After this calibration, the P and Sn peak-to-peak height ratios, with respect to Fe, show scatter of only 15%. In addition to this reduction in scatter, the mean values for the ratios can be corrected by up to a factor of 2. This work shows that the reference materials and methodology are suitable for relating the segregation spectra for P and Sn in low alloy steels from laboratory to laboratory on an absolute and traceable basis.MST/1684b
Source: openalex_inverted_index · Quality: baseline

Research context

OpenAlex topics: Electron and X-Ray Spectroscopy Techniques · Non-Destructive Testing Techniques · Corrosion Behavior and Inhibition

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (2)

Show source-provided references
1. 10.1002/sia.74017120610.1002/sia.740171206 ↗
2. 10.1002/sia.74015120810.1002/sia.740151208 ↗