Evaluating the validity of lightweight talar replacement designs: rational models and topologically optimized models

Yeokyung Kang, Seongjin Kim, Jungsung Kim, Jin Woo Lee, Jong-Chul Park

Biomaterials Research· Volume 26· Issue 1

Abstract

Abstract Background Total talar replacement is normally stable and satisfactory. We studied a rational scaffold talus model for each size range created through topology optimization (TO) and comparatively evaluated a topologically optimized scaffold bone talus model using a finite element analysis (FEA). We hypothesized that the rational scaffold would be more effective for application to the actual model than the topologically optimized scaffold. Methods Size specification for the rational model was performed via TO and inner scaffold simplification. The load condition for worst-case selection reflected the peak point according to the ground reaction force tendency, and the load directions “plantar 10°” (P10), “dorsi 5°” (D5), and “dorsi 10°” (D10) were applied to select worst-case scenarios among the P10, D5, and D10 positions (total nine ranges) of respective size specifications. FEA was performed on each representative specification-standard model, reflecting a load of 5340 N. Among the small bone models selected as the worst-case, an arbitrary size was selected, and the validity of the standard model was evaluated. The standard model was applied to the rational structure during validity evaluation, and the TO model reflecting the internal structure derived by the TO of the arbitrary model was implemented. Result In worst-case selection, the highest peak von Mises stress (PVMS) was calculated from the minimum D5 model (532.11 MPa). Thereafter, FEA revealed peak von Mises stress levels of 218.01 MPa and 565.35 MPa in the rational and topologically optimized models, respectively, confirming that the rational model yielded lower peak von Mises stress. The weight of the minimum model was reduced from 1106 g to 965.4 g after weight reduction through rational scaffold application. Conclusion The rational inner-scaffold-design method is safer than topologically optimized scaffold design, and three types of rational scaffold, according to each size range, confirmed that all sizes of the talus within the anatomical dimension could be covered, which was a valid result in the total talar replacement design. Accordingly, we conclude that an implant design meeting the clinical design requirements, including patient customization, weight reduction, and mechanical stability, should be possible by applying a rational inner scaffold without performing TO design. The scaffold model weight was lower than that of the solid model, and the safety was also verified through FEA.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Foot and Ankle Surgery · Diabetic Foot Ulcer Assessment and Management · Prosthetics and Rehabilitation Robotics

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Central Research & Development Center, Corentec Company, Limited, Seoul, Republic of Korea.

Cellbiocontrol Laboratory, Department of Medical Engineering, Yonsei University College of Medicine, Seoul, Republic of Korea.

Department of Orthopaedic Surgery, Yonsei University College of Medicine, Seoul, Republic of Korea.

References (25)

Show source-provided references
1. 10.5312/wjo.v8.i3.22110.5312/wjo.v8.i3.221 ↗
2. 10.3113/FAI.2007.093310.3113/FAI.2007.0933 ↗
3. 10.1016/j.ocl.2004.02.01010.1016/j.ocl.2004.02.010 ↗
4. 10.1177/107110072094846110.1177/1071100720948461 ↗
5. Kim BC, Choi SJ, Kim HT, Yoo CI, Jung SW, Eun IS, et al. Long term follow-up of avascular necrosis after talar fracture and dislocation: 5 cases. J Korean Foot Ankle Soc. 2005;9:31–37.
6. Choi GW, Choi WJ, Lee JW. Total ankle replacement. J Korean Foot Ankle Soc. 2011;15:132–138.
7. 10.1177/193864001876256710.1177/1938640018762567 ↗
8. 10.1016/j.procir.2017.04.03210.1016/j.procir.2017.04.032 ↗
9. Rishin J, Akoh CC, Chen J, Sharma A, Parekh SG. 3D printed total talus replacement for avascular necrosis of the talus. Foot Ankle Int. 2020;41:1–8.
10. 10.1002/ca.2204310.1002/ca.22043 ↗
11. 10.1097/MD.000000000001714210.1097/MD.0000000000017142 ↗
12. 10.7736/KSPE.2018.35.1.2710.7736/KSPE.2018.35.1.27 ↗
13. 10.11637/kjpa.2012.25.4.18510.11637/kjpa.2012.25.4.185 ↗
14. 10.14193/jkfas.2016.20.4.16310.14193/jkfas.2016.20.4.163 ↗
15. 10.1111/os.1225810.1111/os.12258 ↗
16. 10.1016/j.simpat.2006.09.00110.1016/j.simpat.2006.09.001 ↗
17. 10.1016/j.jbiomech.2014.12.03410.1016/j.jbiomech.2014.12.034 ↗
18. Alamdari A Krovi VN. Chapter two - a review of computational musculoskeletal analysis of human lower extremities. Hum Model Bio-Inspired Robot. 2017:37–73. https://doi.org/10.1016/B978-0-12-803137-7.00003-3.10.1016/B978-0-12-803137-7.00003-3 ↗
19. Silva LM, Stergiou N. The basics of gait analysis. Biomechanics Gait Analy. 2020;164:231.
20. Nandy A, Chakraborty S, Chakraborty J, Venture G. Modern methods for affordable clinical gait analysis: theories and applications in healthcare systems. Acad Press. 2021;ISBN-13:978–0323852456.
21. 10.1371/journal.pone.015443510.1371/journal.pone.0154435 ↗
22. Vidosic JP. Machine design projects. New York: Ronald Press Co; 1957.
23. Ahearne E, Baron S, Keaveney S, Byrne G. An assessment of medical grade cobalt chromium alloy ASTM F1537 as a “Difficult-to- cut (DTC)” material. Proceedings of the Machine Tool Technologies Research Foundation. San Fransisco: Annual meeting; 2015.
24. International Organization for Standardization. ISO 7206-6 Implant for surgery - Partial and total hip joint prostheses - Part 6: Endurance properties testing and performance requirements of neck region of stemmed femoral components. 2013. https://www.iso.org/standard/51186.html. Accessed 07 Feb 2022.
25. 10.9718/JBER.2013.34.1.1410.9718/JBER.2013.34.1.14 ↗