journal article·2013-11

3D-modelling of temperature gradients induced by electrical power dissipation in a 3-body Domen-type calorimeter for absorbed dose measurements

Cristina-Elena Ionita, Daniel Radu, Iordana Astefanoaei

Materials Science and Engineering: B· Volume 178· Issue 19
Publisher page

Abstract

Abstract The effect of thermal gradients that appear in the core of a Domen-type absorbed-dose calorimeter during an electrical calibration run is revised using a finite element 3D-numerical analysis. The local thermal modelling, that is the method allowing for spatio-temporal distribution of temperature within the three bodies of the graphite calorimeter, combined with both a careful analysis on what really happens when a certain amount of electrical power is dissipated in a NTC heating thermistor and the consideration of heat loss through radiation from the surface of the calorimeter bodies, has led to results which are in agreement with those obtained from experimental measurements and reported in the literature.
Source: openaire · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Thermal and Kinetic Analysis · Semiconductor materials and devices · Luminescence Properties of Advanced Materials

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (20)

Show source-provided references
1. Laughlin · 1966 · Vol. II
2. Petree · National Bureau of Standards (United States) · 1962 · 163 · 36
3. Engelke · PTB-Mitteilungen · 1971 · 81 · 185
4. Daures · Physics in Medicine and Biology · 2005 · 50 · 403510.1088/0031-9155/50/17/008 ↗
5. Domen · 1987 · Vol. II
6. Witzani · Metrologia · 1984 · 20 · 7310.1088/0026-1394/20/3/001 ↗
7. McEwen, M.R., Duane, S. Development of a portable graphite calorimeter for photons and electrons, CIRM, 42, pp. 52–64 (1999).
8. Domen · Journal of Research of the NBS, A: Physics and Chemistry · 1974 · 78A · 59510.6028/jres.078A.037 ↗
9. DuSautoy, A.R. The UK primary standard calorimeter for absorbed dose measurement. NPL Report RSA (EXT), 25, Nov (1991)
10. Duane, S. The NPL Primary Standard of Photon Absorbed Dose, CCRI(I)05–35, 3 pp. (2005).
11. Picard Susanne, Burns, D.T., Roger, P. Towards an absorbed dose calorimeter at the BIPM: an experimental method to measure the specific heat capacity of graphite. Progress Report CCRI(I)/05-30, 22 April 2005, 34 pp.
12. Picard Susanne, Burns, D.T., Roger, P. Measurement of the specific heat capacity of graphite, Rapport BIPM-2006/0, 31 pp. (2006).
13. Sapoff · Proceedings of the IEEE · 1963 · 129210.1109/PROC.1963.2560 ↗
14. Paulsen, A., Widera, R., Nerb, H., van Rhee, L. (1986). The CBNM graphite absorbed-dose calorimeter, GE/R/VG/52/86 (private communication).
15. Tietze · 1965
16. Hilbert · 1999
17. Janssens · Metrologia · 1986 · 22 · 26510.1088/0026-1394/22/4/005 ↗
18. Ioniţă · Acta Physica Polonica A · 2010 · 118 · 33310.12693/APhysPolA.118.550 ↗
19. Ioniţă · Biomaterials in Biophysics Medical Physics and Ecology · 2009 · 6
20. Ionita · Romanian Journal of Physics · 2013 · 58 · 220