Nano-Mechanical Transduction of Polymer Micro-Cantilevers to Detect Bio-Molecular Interactions
Source abstract
Abstract
Using variothermal polymer micro-injection molding, disposable arrays of eight polymer micro-cantilevers each 500 μm long, 100 μm wide and 25 μm thick were fabricated. The present study took advantage of an easy flow grade polypropylene. After gold coating for optical read-out and asymmetrical sensitization, the arrays were introduced into the Cantisens® Research system to perform mechanical and functional testing. We demonstrate that polypropylene cantilevers can be used as biosensors for medical purposes in the same manner as the established silicon ones to detect single-stranded DNA sequences and metal ions in real-time. A differential signal of 7 nm was detected for the hybridization of 1 μM complementary DNA sequences. For 100 nM copper ions the differential signal was found to be (36 ± 5) nm. Nano-mechanical sensing of medically relevant, nanometer-size species is essential for fast and efficient diagnosis.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Mechanical and Optical Resonators · Force Microscopy Techniques and Applications · Carbon Nanotubes in Composites
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
University of Basel 1 Biomaterials Science Center, , c/o University Hospital, Basel, 4031, Switzerland
Paul Scherrer Institut 2 Laboratory for Micro- and Nanotechnology, , Villigen PSI, 5232, Switzerland
University of Applied Sciences Northwestern Switzerland 3 Institute of Chemistry and Bioanalytics, , Muttenz, 4132, Switzerland
University of Applied Sciences Northwestern Switzerland 4 Institute of Polymer Nanotechnology, , Windisch, 5210, Switzerland
References (20)
Show source-provided references
1. Science · 2000 · 288 · 31610.1126/science.288.5464.316 ↗
2. PNAS · 2002 · 99 · 978310.1073/pnas.152330199 ↗
3. Nat Nanotechnol · 2006 · 1 · 21410.1038/nnano.2006.134 ↗
4. Ultramicroscopy · 2000 · 82 · 110.1016/S0304-3991(99)00123-0 ↗
5. Sens Actuators · 2001 · B 77 · 122
6. Eur J Nanomed · 2008 · 1 · 4410.1515/EJNM.2008.1.1.44 ↗
7. Can Med Assoc J · 1998 · 138 · 23
8. Proc Royal Soc London Ser A · 1909 · 82 · 17210.1098/rspa.1909.0021 ↗
9. J Phys · 2007 · 61 · 61210.1088/1742-6596/61/1/123 ↗
10. Methods · 2005 · 37 · 5710.1016/j.ymeth.2005.05.011 ↗
11. Anal Bioanal Chem · 2004 · 379 · 946
12. Sensors · 2008 · 8 · 660510.3390/s8106605 ↗
13. BMC Biotechnol · 2011 · 11 · 12
14. Toxicol Appl Pharmacol · 1996 · 140 · 110.1006/taap.1996.0191 ↗
15. Phys Chem Chem Phys · 2006 · 8 · 2010.1039/b511146c ↗
16. Science · 1997 · 276 · 202110.1126/science.276.5321.2021 ↗
17. J Micromech Microeng · 1993 · 3 · 410.1088/0960-1317/3/1/002 ↗
18. Langmuir · 2008 · 24 · 870710.1021/la800287e ↗
19. Meas Sci Technol · 2006 · 17 · 296410.1088/0957-0233/17/11/015 ↗
20. PLoS ONE · 2008 · 3 · e361010.1371/journal.pone.0003610 ↗