journal article·2013Open access

A PM1.0/2.5/10 Trichotomous Virtual Impactor Based Sampler: Design and Applied to Arid Southwest Aerosols Part I: Design

Virgil Marple, Dale Lundgren, Bernard Olson

KONA Powder and Particle Journal· Volume 30· Issue 0

Abstract

A PM1.0/2.5/10 Trichotomous sampler has been developed to determine if the particles in the saddle point between the coarse and fine particle modes (specifically the 1.0 μm to 2.5 μm size range) are primarily coarse or fine mode particles. The sampler consists of a standard high volume sampler with two high volume virtual impactors (one with a cut size of 2.5 μm and the other with a cut size of 1.0 μm) inserted between the PM10 inlet and the 8×10 inch (20×25 cm) after filter. By using nine 47 mm filters, at various locations within the trichotomous sampler, a technique has been developed for subtracting out the effects of particles smaller than the cutsize of a virtual impactor from the minor flow particle collection.
Source: openalex_inverted_index · Quality: baseline

Research context

OpenAlex topics: Indoor Air Quality and Microbial Exposure · Air Quality and Health Impacts · Wind and Air Flow Studies

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Particle Calibration Laboratory, University of Minnesota

University of Florida

Funding

National Mining Association

References (13)

Show source-provided references
1. Berglund, R. N. and Liu, B. Y. H. (1973): Generation of monodisperse aerosol standards, Environmental Science Technology, Vol.7, pp.147–153.10.1021/es60074a001 ↗
2. Chen, B. T., Yeh, H. C. and Cheng, Y. S. (1986): Performance of a modified virtual impactor, Aerosol Sci. Technol. Vol.5, pp.369–376.10.1080/02786828608959101 ↗
3. John, W. (2011): “Size distribution characteristics of aerosols”, Aerosol Measurement, 3rd edition, eds: Kulkarni, Baron and Willeke, pp.41–54.10.1002/9781118001684.ch4 ↗
4. Marple, V. A. (1970): “A Fundamental Study of Inertial Impactors”, Ph.D. dissertation, University of Minnesota.10.2172/4095434 ↗
5. Marple, V. A. (2004): History of Impactors-The first 110 Years, Aerosol Science and Technology, Vol.38, pp.247–292.10.1080/02786820490424347 ↗
6. Marple, V. A. and Chein, C. M. (1980): Virtual Impactors: A Theoretical Study, Environ. Sci. Technology, Vol.14, pp.976–985.10.1021/es60168a019 ↗
7. Marple, V. A., Liu, B. Y. H. and Burton, R. M. (1990): High-volume virtual impactor for sampling fine and coarse particles, J. Air Waste Management Assoc. Vol.40, pp.762–767.10.1080/10473289.1990.10466722 ↗
8. Marple, V. A. and Olson, B. A. (1995): “A high volume PM10/2.5/1.0 trichotomous sampler”, In Particulate Matter: Health and Regulatory Issues, VIP-49, Air and Waste Management Assoc. International Specialty Conference, Pittsburg, PA, pp.237–261.
9. Masuda, H., Hochrainer, D., and Stober, W. (1978): An improved virtual impactor for particle classification and generation of test aerosols with narrow size distributions, J. Aerosol Sci. Vol.10, pp.275–287.10.1016/0021-8502(79)90043-0 ↗
10. McFarland, A. R., Ortiz, C. A. and Bertch, Jr.R. W. (1984): A 10 μm Outpoint size selective inlet for Hi-Vol Samplers, J. Air Pollution Control Assoc. Vol.34, pp.544–547.10.1080/00022470.1984.10465773 ↗
11. U.S. Environmental Protection Agency (1982): “Air quality criteria for particulate matter and sulfur”, EPA-600/882-029b, December, 1982.
12. Whitby, K. T. (1978): The physical characteristics of sulfur aerosols, Atmospheric Environment. Vol.12, pp.135–159.10.1016/B978-0-08-022932-4.50018-5 ↗
13. Xu, X, (1991): “A Study of Virtual Impactors”, Ph.D. dissertation, University of Minnesota.