journal article·2017-04Open access

Tessellated permanent magnet circuits for flow-through, open gradient separations of weakly magnetic materials

Lee R. Moore, P. Stephen Williams, Jeffrey J. Chalmers, Maciej Zborowski

Abstract

Emerging microfluidic-based cell assays favor label-free red blood cell (RBC) depletion. Magnetic separation of RBC is possible because of the paramagnetism of deoxygenated hemoglobin but the process is slow for open-gradient field configurations. In order to increase the throughput, periodic arrangements of the unit magnets were considered, consisting of commercially available Nd-Fe-B permanent magnets and soft steel flux return pieces. The magnet design is uniquely suitable for multiplexing by magnet tessellation, here meaning the tiling of the magnet assembly cross-sectional plane by periodic repetition of the magnet and the flow channel shapes. The periodic pattern of magnet magnetizations allows a reduction of the magnetic material per channel with minimal distortion of the field cylindrical symmetry inside the magnet apertures. A number of such magnet patterns are investigated for separator performance, size and economy with the goal of designing an open-gradient magnetic separator capable of reducing the RBC number concentration a hundred-fold in 1 mL whole blood per hour.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Microfluidic and Bio-sensing Technologies · Characterization and Applications of Magnetic Nanoparticles · Lattice Boltzmann Simulation Studies

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (20)

Show source-provided references
1. Rare cell separation and analysis by magnetic sorting10.1021/ac200550d ↗
2. Small but mighty: how the MACS®-technology based on nanosized superparamagnetic particles has helped to analyze the immune system within the last 20 years10.1002/cyto.a.20918 ↗
3. The biocompatibility and toxicity of magnetic particles
4. Nanoparticles affect PCR primarily via surface interactions with PCR components: using amino-modified silica-coated magnetic nanoparticles as a main mode10.1021/am508842v ↗
5. Quantification of non-specific binding of magnetic micro- and nanoparticles using cell tracking velocimetry: implication for magnetic cell separation and detection10.1002/bit.22635 ↗
6. Open gradient magnetic red blood cell sorter evaluation on model cell mixtures10.1109/TMAG.2012.2225098 ↗
7. Red blood cell magnetophoresis10.1016/S0006-3495(03)75069-3 ↗
8. Shapiro · 2005
9. Microfluidic, marker-free isolation of circulating tumor cells from blood samples10.1038/nprot.2014.044 ↗
10. Negative selection of hematopoietic progenitor cells by continuous magnetophoresis10.1016/j.exphem.2006.12.009 ↗
11. Magnetophoretic cell sorting is a function of antibody binding capacity10.1021/bp020285e ↗
12. Flow rate optimization for the quadrupole magnetic cell sorter10.1021/ac990284+ ↗
13. High gradient magnetic separation of red cells from whole blood10.1109/TMAG.1975.1058970 ↗
14. Distribution of erythrocyte in a model vessel exposed to inhomogeneous magnetic fields
15. The magnetic properties and structure of hemoglobin, oxyhemoglobin and carbonmonoxyhemoglobin10.1073/pnas.22.4.210 ↗
16. Erythrocyte enrichment in hematopoietic progenitor cell cultures based on magnetic susceptibility of the hemoglobin10.1371/journal.pone.0039491 ↗
17. Continuous magnetic separation of blood components from whole blood10.1109/77.828383 ↗
18. Separation of red blood cells by magnetic means10.1016/S0304-8853(00)00479-0 ↗
19. Three-dimensional modeling of a portable medical device for magnetic separation of particles from biological fluids10.1088/0031-9155/52/17/007 ↗