Calibration of a permeameter for measuring soft magnetic materials using long cylindrical samples characterized by demagnetizing-corrected solenoid method
Source abstract
Abstract
Abstract A technique is developed to calibrate the performance of a permeameter for measuring soft magnetic materials, having a C-shaped magnetic potential H-coil near the sample surface and a B-coil surrounding the sample, by using cylindrical standard samples. The true M ( H ) curve for the standard samples of large length-to-diameter ratios is obtained by the demagnetizing-corrected solenoid method recently developed.
Source: openaire · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Magnetic Properties and Applications · Magnetic properties of thin films · Magneto-Optical Properties and Applications
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (12)
Show source-provided references
1. Direct-current magnetic measurements for soft magnetic materials, ASTM 1970.
2. ASTMA341/A341M Standard Test Method for Direct Current Magnetic Properties of Materials Using D-C Permeameters and the Ballistic Test Methods.
3. International Standard: IEC-60404-4 Magnetic Materials – Part 4: Methods of measurement of d.c. magnetic properties of magnetically soft materials, 2008.
4. Harris · 1952
5. Cullity · 2009
6. Chen · J. Magn. Magn. Mater. · 2018 · 449 · 44710.1016/j.jmmm.2017.10.069 ↗
7. REMAGRAPH C – 500 for computer controlled measurement of the hysteresis curves of soft magnetic materials.www.magnet-physik.de.
8. Manual of FE-2100SD DC hysteresis curves test system for magnetic materialshttp://m.instrument.com.cn/bbs/d-6236870.html.
9. Chen · IEEE Trans. Magn. · 1991 · 27 · 360110.1109/20.102932 ↗
10. Chen · J. Magn. Magn. Mater. · 2006 · 306 · 13510.1016/j.jmmm.2006.02.235 ↗
11. Zhu · Metallic Funct Mater · 2016 · 23 · 1
12. D.-X. Chen, Ballistic and Bridge Methods of Magnetic Measurements of Materials, China Metrology Publication, 1990.