Contributing factors to fires in clinical settings during medical laser applications
Source abstract
Abstract
An estimated 650 fires occur in the surgical environment annually in the U.S.; a figure which may be severely underreported. Any fire in the surgical environment can have devastating implications for patient safety and is clearly an occupational risk to healthcare personnel. Literature searches of the PubMed and Rockwell Laser Industries Laser Accident Databases were performed in order to (1) identify factors and sources affecting the risk of fire in the medical environment, namely, oxidizers, ignition sources, and fuels, and (2) to identify areas of future research for prevention. All papers relevant to the flammability of commonly used medical supplies when subjected to laser radiation, including clinical case reports and publicly available incident data, were identified, reviewed, and summarized. There are numerous materials that are ubiquitous within the medical setting that can ignite when accidentally contacted with laser radiation, including gauze, drapes, sponges, adhesive tapes, and gowns, skin preparatory solutions and ointments, and endotracheal tubes. Future research is warranted to determine the dependence of the ignition potential of the various fuels on laser type and operational parameters. The purpose of this analysis is to summarize and present all of the seminal published literature pertaining to medical laser-related fires and their prevention in order to understand the contributing factors and extent to which these hazards exist, to inform the medical laser community about these hazards, and to prioritize future areas of occupational health and safety research.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Ocular and Laser Science Research · Traumatic Ocular and Foreign Body Injuries · Radiation Dose and Imaging
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
University of Illinois at Chicago Environmental and Occupational Health Sciences, , 2121 W. Taylor Street—MC 922, Chicago, Illinois 60612
References (55)
Show source-provided references
1. 2010
2. Fire safety in the operating room10.1097/ACO.0b013e328318693a ↗
3. Laser ignition of surgical drape materials in air, 50% oxygen, and 95% oxygen10.1097/00000542-200405000-00019 ↗
4. Dangers from OR fires still a serious problem10.1023/A:1009908200365 ↗
5. Fires in the operating room and intensive care unit: Awareness is the key to prevention10.1213/01.ane.0000183641.09008.f2 ↗
6. Fire in the operating room during tracheotomy: A case report
7. Fires and burns during plastic surgery10.1097/00000637-200101000-00016 ↗
8. A clinician’s guide to surgical fires. How they occur, how to prevent them, how to put them out
9. Fire in the operating room during open heart surgery: A case report
10. Oxygen leakage around the laryngeal mask airway during laser treatment of port-wine stains in children10.1213/00000539-199403000-00012 ↗
11. Fire in the operating room during tracheostomy10.1097/00007611-199803000-00005 ↗
12. Remote fire with the pulsed dye laser: Risk and prevention10.1016/S0190-9622(96)90457-0 ↗
13. CO2 laser safety considerations in facial skin resurfacing10.1097/00006534-199710000-00034 ↗
14. Principles of CO2/erbium laser safety10.1046/j.1524-4725.2000.09187.x ↗
15. Ignition potential of the 585-nm pulsed-dye laser. Review of the literature and safety recommendations10.1016/1076-0512(96)00279-8 ↗
16. Anesthesiologists must do a better job of preventing operating room fires10.1016/j.jclinane.2005.12.003 ↗
17. See supplementary material at http://dx.doi.org/10.2351/1.4704850 Table 1. Summary of Flammability Studies on Medical Supplies and Equipment.
18. Incendiary potential of the flash-lamp pumped 585-nm tunable dye laser10.1213/00000539-199008000-00009 ↗
19. Laser ignition of surgical drapes
20. Laser safety features of eye shields10.1002/(SICI)1096-9101(1996)18:3<309::AID-LSM13>3.0.CO;2-0 ↗
21. Hair ignition by dye laser for port-wine stain: Risk factors evaluated10.1080/14628839950516986 ↗
22. Possible hazards of cryogen use with pulsed dye laser. A case report and summary10.1046/j.1524-4725.1999.08198.x ↗
23. CO2 laser resistance of various ointments and tapes10.1097/00000542-199105000-00046 ↗
24. Skin preparation for laser therapy
25. Endotracheal tube ignition during laser surgery of the larynx10.1001/archotol.1979.00790210059013 ↗
26. Intratracheal fire ignited by the nd-YAG laser during treatment of tracheal stenosis10.1378/chest.84.3.295 ↗
27. Endotracheal tube ignition during CO2 laser surgery of the larynx
28. Carbon dioxide laser induced airway fire during larynx surgery: Case report
29. Laser-induced endotracheal tube fire10.1097/00000542-198111000-00018 ↗
30. Laser initiated endotracheal tube explosion
31. Ignition of an endotracheal tube during laser microsurgery10.1097/00000542-198008000-00015 ↗
32. Indirect ignition of the endotracheal tube during carbon dioxide laser surgery10.1001/archotol.1980.00790340047012 ↗
33. Long-term follow-up after laser-induced endotracheal fire10.1258/0022215021910366 ↗
34. Endotracheal tube fires during carbon dioxide laser surgery on the larynx–a case report
35. Complications of CO2 laser surgery of the larynx
36. Airway ignition during CO2 laser laryngeal surgery and high frequency jet ventilation10.1097/00003643-200003000-00013 ↗
37. Fire hazard during CO2 laser microsurgery on the larynx and trachea10.1213/00000539-197601000-00039 ↗
38. Tracheostomy tube ignition during microlaryngeal surgery using diode laser: A case report10.1016/S1607-551X(09)70308-0 ↗
39. Value of endotracheal tube safety in laryngeal laser surgery10.1002/lsm.10174 ↗
40. Flammability of endotracheal tubes during nd-YAG laser application in the airway10.1097/00000542-198611000-00010 ↗
41. Fires of endotracheal tubes of three different materials during carbon dioxide laser surgery
42. Prevention of fire hazards associated with use of carbon dioxide lasers10.1213/00000539-198112000-00008 ↗
43. Acute management of laser-ignited intratracheal explosion10.1288/00005537-198109000-00002 ↗
44. Fatal inhalation injury caused by airway fire during tracheostomy10.1111/j.1399-6576.2007.01280.x ↗
45. Comparative analysis of the safety of endotracheal tubes with the KTP laser10.1288/00005537-198907000-00017 ↗
46. What is the safest endotracheal tube for nd-YAG laser surgery? A comparative study10.1213/00000539-198912000-00020 ↗
47. Which is the safest endotracheal tube for use with the CO2 laser? A comparative study10.1016/0952-8180(92)90069-D ↗
48. Effects of the holmium:YAG and erbium:YAG lasers on endotracheal tubes10.1159/000276391 ↗
49. Possible hazards from irradiation with the carbon dioxide laser10.1002/lsm.1900060117 ↗
50. Lasers in medicine and surgery10.1001/jama.1986.03380070106029 ↗
51. High incidence of CO2 laser beam contact with the tracheal tube during operations on the upper airway10.1016/0952-8180(89)90075-5 ↗
52. Evaluation of five metallic tapes for protection of endotracheal tubes during CO2 laser surgery10.1213/00000539-198903000-00039 ↗
53. What is the safest foil tape for endotracheal tube protection during nd-YAG laser surgery? A comparative study10.1097/00000542-199003000-00025 ↗
54. Evaluation of foil coverings for protecting plastic endotracheal tubes from the potassium-titanyl-phosphate laser
55. Evaluation of a new laser-resistant fabric and copper foil-wrapped endotracheal tube10.1097/00005537-199607000-00011 ↗