A sub-cc nonlinear piezoelectric energy harvester for powering leadless pacemakers
Source abstract
Abstract
A miniature nonlinear piezoelectric energy harvester is developed to power state of the art leadless cardiac pacemakers from cardiac motions. The energy harvester is integrated in the leadless pacemaker and is connected to the myocardium. The energy harvester converts myocardial motions to electricity to power leadless pacemakers. The energy is stored in a battery or supercapacitor and is used for pacing. The device is composed of a bimorph piezoelectric beam confined in a gray iron frame. The system is assembled at high temperature and operated at the body temperature. The mismatch in the coefficients of thermal expansion of the beam and the frame causes the beam to buckle in body temperature. This intentional buckling makes the beam unstable and improves the power production and robustness of the device. Having high natural frequency is a major problem in microelectromechanical systems energy harvesters. Considering the small size of the energy harvester, [Formula: see text], the natural frequency is expected to be high. In our design, the natural frequency is lowered significantly using a buckled beam and a proof mass. Since the beam is buckled, the design is bistable and nonlinear, which could increase the output power. In this article, the device is analytically modeled, and the natural frequencies and mode shapes of the energy harvester are analytically derived. The terms corresponding to geometric nonlinearities are included in the electromechanical coupled governing equations. The simulations show that the device generates sufficient electricity to power leadless pacemakers.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Innovative Energy Harvesting Technologies · Advanced Sensor and Energy Harvesting Materials · Wireless Power Transfer Systems
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Department of Mechanical and Aerospace Engineering, University at Buffalo, The State University of New York (SUNY), Buffalo, NY, USA
References (33)
Show source-provided references
1. 10.1088/0964-1726/24/11/11500510.1088/0964-1726/24/11/115005 ↗
2. 10.1115/DETC2016-6047610.1115/DETC2016-60476 ↗
3. 10.1063/1.494288210.1063/1.4942882 ↗
4. Bilitch M (1981) Leadless cardiac pacer. Google Patents.
5. 10.1016/0022-460X(88)90397-510.1016/0022-460X(88)90397-5 ↗
6. 10.1109/GreenCom.2012.15110.1109/GreenCom.2012.151 ↗
7. 10.1088/0964-1726/21/3/03502110.1088/0964-1726/21/3/035021 ↗
8. 10.1073/pnas.131723311110.1073/pnas.1317233111 ↗
9. Deterre M · 2013
10. Emam SA · 2002
11. 10.1016/j.compstruct.2012.12.04410.1016/j.compstruct.2012.12.044 ↗
12. 10.1016/j.compstruct.2008.06.00610.1016/j.compstruct.2008.06.006 ↗
13. 10.1007/s00707-013-0941-z10.1007/s00707-013-0941-z ↗
14. 10.1007/BF0251331510.1007/BF02513315 ↗
15. 10.1002/adma.20140056210.1002/adma.201400562 ↗
16. 10.1109/58.53548010.1109/58.535480 ↗
17. 10.1063/1.367910210.1063/1.3679102 ↗
18. 10.1063/1.359527810.1063/1.3595278 ↗
19. 10.1557/mrs.2012.27510.1557/mrs.2012.275 ↗
20. 10.2514/1.2472010.2514/1.24720 ↗
21. 10.2514/3.1053110.2514/3.10531 ↗
22. 10.1115/1.400278610.1115/1.4002786 ↗
23. Nippoldt D, Whiting J (2014a) Micra transcatheter pacing system: device volume characterization comparison (Medtronic data on file).
24. Nippoldt D, Whiting J (2014b) Micra transcatheter pacing system: FIH announcement (Medtronic data on file).
25. 10.1016/S0263-8223(03)00146-610.1016/S0263-8223(03)00146-6 ↗
26. 10.1088/0964-1726/23/7/07500310.1088/0964-1726/23/7/075003 ↗
27. Shaker FJ · 1975
28. 10.1007/978-1-4614-5705-3_210.1007/978-1-4614-5705-3_2 ↗
29. Williams E, Jon W (2014) Micra transcatheter pacing system: size comparison (Medtronic data on file).
30. 10.1088/1742-6596/660/1/01201310.1088/1742-6596/660/1/012013 ↗
31. 10.1109/TUFFC.90310.1109/TUFFC.903 ↗
32. 10.1063/1.481651810.1063/1.4816518 ↗
33. 10.1007/s10439-012-0623-310.1007/s10439-012-0623-3 ↗