Design, analysis, and fabrication of a piezoelectric force tray for total knee replacements
Source abstract
Abstract
Force plates have been widely adopted in biomechanical gait analysis to measure reaction forces and the center of pressure. In this work, the force plate concept is miniaturized and extended for use within the polyethylene bearing insert of a total knee replacement. A simplified rectangular-shaped force plate with multiple integrated piezoelectric sensors, including designs with six and eight transducers, is presented in this work. The performance of the sensory system is investigated through finite element analysis and experimental validation. Initially, the ability of the two designs in sensing compartmental forces and contact point locations on one side of the force plate is numerically investigated. Selected designs of the force plate are then fabricated and used to experimentally validate the performance of the system. The results show a maximum error of less than 6% and 4.5% in compartmental force amplitude sensing for the force plates with six and eight transducers, respectively. The force plates were able to detect the contact point location with maximum errors of less than 1 mm. The relatively small sensing error quantities show the potential of using a piezoelectric force plate sensor design in total knee replacement as well as other force sensing applications.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Advanced Sensor and Energy Harvesting Materials · Innovative Energy Harvesting Technologies · Dielectric materials and actuators
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Department of Mechanical Engineering, Tennessee Tech University, Cookeville, TN, USA
Funding
National Institute of Arthritis and Musculoskeletal and Skin Diseases · national institutes of health
References (29)
Show source-provided references
1. 10.1109/TBME.2010.205880610.1109/TBME.2010.2058806 ↗
2. 10.1088/0964-1726/16/3/R0110.1088/0964-1726/16/3/R01 ↗
3. APC International Ltd (2018) Physical and piezoelectric properties of APC materials. Available at: https://www.americanpiezo.com/apc-materials/physical-piezoelectric-properties.html (accessed 11 November 2018).
4. 10.1016/0021-9290(90)90169-410.1016/0021-9290(90)90169-4 ↗
5. Curie J · Comptes rendus de l’Académie des Sciences · 1881 · 93 · 1137
6. 10.1016/j.jbiomech.2004.02.01110.1016/j.jbiomech.2004.02.011 ↗
7. Hubbard B · Journal of Otolaryngology-Head and Neck Surgery · 2012 · 41 · S31
8. 10.1088/1361-665X/aaf3f110.1088/1361-665X/aaf3f1 ↗
9. Jaffe B · 1971
10. Lippmann GJ · Annales de Chimie et de Physique · 1881 · 24 · 145
11. 10.1016/j.jbiomech.2015.12.01610.1016/j.jbiomech.2015.12.016 ↗
12. 10.1016/S0268-0033(98)00079-510.1016/S0268-0033(98)00079-5 ↗
13. 10.2106/00004623-200100021-0001310.2106/00004623-200100021-00013 ↗
14. 10.1080/14763141.2016.121387510.1080/14763141.2016.1213875 ↗
15. 10.1109/TMECH.2005.84470410.1109/TMECH.2005.844704 ↗
16. 10.1109/TMECH.2005.85248210.1109/TMECH.2005.852482 ↗
17. Polymaker (2018) PolyLite PLA (formerly PolyPlus PLA). Available at: https://us.polymaker.com/product/polylite-pla/ (accessed 17 June 2018).
18. 10.1016/j.jmbbm.2018.12.01810.1016/j.jmbbm.2018.12.018 ↗
19. 10.2106/00004623-200302000-0001210.2106/00004623-200302000-00012 ↗
20. 10.1088/1361-665X/aa814e10.1088/1361-665X/aa814e ↗
21. 10.1109/TMECH.2018.279418210.1109/TMECH.2018.2794182 ↗
22. 10.1088/1361-665X/aad75510.1088/1361-665X/aad755 ↗
23. 10.1016/S0021-9290(99)00109-810.1016/S0021-9290(99)00109-8 ↗
24. 10.1115/1.128968710.1115/1.1289687 ↗
25. 10.1007/0-387-33531-510.1007/0-387-33531-5 ↗
26. 10.1016/j.arth.2013.07.02410.1016/j.arth.2013.07.024 ↗
27. 10.1016/S0883-5403(98)90009-210.1016/S0883-5403(98)90009-2 ↗
28. Voigt W · 1910
29. Wong C · IEEE Sensors Journal · 2015 · 15 · 2747