article·2011-01-01

Performance Improvement of the NiTi Alloy Wire Used for Aortic Stent-graft

Yuan Zhen

Abstract

In order to improve the physical properties and biocompatibility of the aortic stent-graft,the effects of heat treatment and electrochemical polishing on nickel-titanium(NiTi) alloy wire during the stent-graft preparation were studied.It is found that heat treatment could help shaping the stent-graft and enhance the mechanical proper-ties of the NiTi wire.The ideal heat-treatment process is heating at 480℃ and 510℃ for 7 min,respectively.By electrochemical polishing,the surface quality and fatigue life of the wire is highly improved.Meanwhile,electro-chemical polishing results in reduction of the possibility of the mural thrombus formation and endothelial hyperplasy.
Source: openalex_inverted_index · Quality: baseline

Research context

OpenAlex topics: Coronary Artery Anomalies · Aortic Disease and Treatment Approaches · Aortic aneurysm repair treatments

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References

Reference metadata was not supplied for this record.